High-energy gamma-ray spectrometry with LYSO:Ce scintillators for fusion plasma experiments.
S Mianowski, P Beaumont, M Beldishevski, Z Ghani, V G Kiptily, Z Mianowska, B Phoenix, A Tilley
UKAEA, Culham Campus, Abingdon, Oxfordshire OX14 3DB, United Kingdom. · School of Physics and Astronomy, The University of Birmingham, Edgbaston B15 2TT, United Kingdom.
Key points
From the paper's conclusion- The results demonstrate that detector segmentation, combined with coincidence techniques and pileup deconvolution, enables efficient high-count-rate gamma-ray spectrometry using SiPM-based detectors and offers a promising solution for fusion plasma diagnostics in high-magnetic-field environments.
Abstract
High-performance gamma-ray spectrometry for fusion plasma diagnostics requires high detection efficiency in the MeV energy range, low pileup losses at high count rates, and an enhanced peak-to-background ratio. These requirements cannot be achieved simultaneously with a single large-volume scintillation detector. To address this limitation, a segmented detector consisting of a stack of three optically isolated lutetium-yttrium oxyorthosilicate doped with cerium scintillation crystals with a silicon photomultiplier (SiPM) readout was developed and tested. Each crystal was equipped with an independent, synchronized readout channel and digitized using a high-speed waveform-acquisition system, enabling operation at count rates of 200 kcps per crystal. Energy was reconstructed using charge integration, whereas event timing was determined using constant-fraction discrimination to enable coincidence measurements. System performance was evaluated using a Ga-66 gamma-ray source that emits multiple lines at energies of up to 5 MeV. A proof-of-principle demonstration of pileup deconvolution was performed using a pattern-based algorithm applied to raw waveforms. The method enabled overlapping events to be identified and reconstructed at a count rate of ∼200 kcps per scintillator. Coincidence selection identified a subset of events corresponding to ∼21% of the 2752 keV full-energy peak counts observed in the individual detector spectra, providing additional information on correlated interactions within the segmented detector. The results demonstrate that detector segmentation, combined with coincidence techniques and pileup deconvolution, enables efficient high-count-rate gamma-ray spectrometry using SiPM-based detectors and offers a promising solution for fusion plasma diagnostics in high-magnetic-field environments.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Gamma-ray response of a GAGG:Ce scintillator coupled to a 2 × 2 SiPM arrayJournal of Instrumentation · 2026
- A Large-Volume GAGG(Ce) Detector Module for In-Beam $\gamma$-Ray SpectroscopyarXiv · 2026
- A SiPM-based plastic scintillator for beam intensity and Time-of-Flight measurements at FRIBJournal of Instrumentation · 2026
- Development of ultrafast and radiation-hard GAGG for the next-generation of High-Energy Physics calorimetersNuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 2026
- Development and High-Count-Rate Performance of Multi-Channel HPGe Spectrometers for X- and Gamma-Ray SpectroscopyInstruments · 2026
- A large neutron/gamma detector based on the EJ-276G plastic scintillator read out with multiple SiPM arraysNuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 2025