Development and High-Count-Rate Performance of Multi-Channel HPGe Spectrometers for X- and Gamma-Ray Spectroscopy

Rais Nurgalejev, Olga Savina, Taras Mykytiuk, Serhii Pohuliai, Maryna Kanapelka, Vladimir Kondratjev, Aleksander Sokolov, Vladimir Gostilo

Key points

From the paper's conclusion
  • The 600 eV target was not consistently met.

Abstract

This paper presents the development and comparative investigation of multi-detector HPGe spectrometers for operation at high count rates, with emphasis on K-series X-ray lines of rare-earth and heavy elements in X-ray fluorescence analysis. Two detector architectures were developed and tested: a six-channel multi-crystal assembly based on individual P-type semi-planar HPGe detectors and a six-strip segmented planar HPGe detector with amorphous-germanium contacts. The segmented detector with cooled FETs demonstrated energy resolutions of 560 eV at 5.9 keV, 700 eV at 59.5 keV, and 1.29 keV at 662 keV at low count rates. At 661.7 keV, the FWHM increased to 3.2–3.5 keV at input count rates of 40–57 kcps per segment. The multi-crystal assembly achieved average FWHM values of 370 eV at 14 keV, 450 eV at 59.5 keV, 560 eV at 122 keV, and 1.16 keV at 662 keV at low count rates. At maximum throughput, FWHM values at 59.5 keV were 580 and 805 eV under 238Pu and 137Cs loading, respectively (ICR 100–120 kcps; OCR 57 and 46 kcps per channel). The 600 eV target was not consistently met. The multi-crystal assembly also recorded preliminary tube-excited K-line XRF spectra of Sm, Ho, and W samples.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Development and High-Count-Rate Performance of Multi-Channel HPGe Spectrometers for X- and Gamma-Ray Spectroscopy — Science Skim