A SiPM-based plastic scintillator for beam intensity and Time-of-Flight measurements at FRIB
Salvatore Di Carlo, Marco Cortesi, Giordano Cerizza, Genie Jhang, Chandana Sumithrarachchi, Christopher Izzo, Xiangcheng Chen
Key points
From the paper's conclusion- The system provides reliable counting and timing measurements at beam intensities exceeding 10 6 pps, well above the limits of previous Si-PIN-based instrumentation, which was restricted to a few kpps.
Abstract
A fast scintillator-based beam counting detector has been developed for high-rate beam monitoring in the stopped and reaccelerated beam area of the Facility for Rare Isotope Beams (FRIB). The detector counts heavy-ion beam particles with energies up to 200 MeV/u on an event-by-event basis and is designed to support transmission efficiency studies, beam tuning, and beam diagnostics. The device consists of a large-area plastic scintillator read out by two arrays of silicon photomultipliers (SiPMs) mounted on custom-designed printed circuit boards. Each board hosts 23 SiPMs, and the two identical boards are mounted on opposite sides of the scintillator to maximize photon collection and to mitigate the effects of position-dependent light absorption in the scintillating medium. The front-end electronics implement a passive summation scheme that combines the signals from multiple SiPMs, extending the dynamic range beyond that of single-threshold discrimination and enabling optimized fast-timing performance at high event rates. The system provides reliable counting and timing measurements at beam intensities exceeding 10 6 pps, well above the limits of previous Si-PIN-based instrumentation, which was restricted to a few kpps. The achieved timing resolution enables atomic mass identification through ToF measurements, with about 8.4 σ separation between neighboring nuclear species. The system has been integrated into the standard beam diagnostics suite at FRIB and is routinely used to support beam development in preparation for user experiments.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The scintillating fibre beam monitor for the CHNet-MAXI experiment at ISIS: characterisation of the light collection and detection systemarXiv · 2026
- High-energy gamma-ray spectrometry with LYSO:Ce scintillators for fusion plasma experiments.The Review of scientific instruments · 2026
- Measurements of the micro-spill structure of medical cyclotron and synchrotron beams and its impact on pulse pileupJournal of Instrumentation · 2026
- Characterization of a Resistively Multiplexed SiPM Readout for Position-Sensitive Scintillating Fibre Radiation Detectors.Sensors (Basel, Switzerland) · 2026
- Neutron irradiation damage on Silicon Photomultipliers and in-situ current annealing studies for the CBM RICH detectorNuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 2026
- Radiation monitoring in the scattered field of a clinical electron FLASH beam with Timepix4Journal of Instrumentation · 2026