Expired endoscopic devices remain sterile and functional: an international multicentre study challenging current shelf-life criteria.
Enrique Rodriguez de Santiago, David Villaescusa Arenas, Elia Gómez G de la Pedrosa, Marina García-Castellanos, João A Cunha Neves, Anastasia Tsalouka, Christoph Römmele, Mathieu Pioche,
Hospital Universitario Ramón y Cajal. Universidad de Alcalá. IRYCIS, Gastroenterology and Hepatology, Spain, Madrid. · Universidad de Alcalá, Universidad de Alcalá, Community of Madrid, Spain, Alcala de Henares.
Key points
From the paper's conclusion- These findings challenge current shelf-life assumptions and support further regulatory evaluation, but should not be interpreted as evidence supporting routine clinical use of expired devices.
Abstract
Background and study aims Expired medical devices represent an environmental and economic concern, yet their magnitude and true shelf life remain poorly defined. We aimed to quantify the volume and cost of expired endoscopic devices and evaluate their sterility and functionality beyond the labelled expiry date. Methods We conducted an investigator-initiated, prospective study (December 2023-May 2025) across 25 endoscopy units from Europe, Canada, and the USA. Expired devices were inventoried during one-week audits. An institutional questionnaire assessed storage practices. Microbiological sterility and ex vivo functionality were assessed in randomly selected samples. Physicomechanical testing included matched case-control comparison with non-expired counterparts. Results We identified 1,448 expired devices (median per centre, 42; range 3-262), with an estimated cost of €192,141.67. Of these, 273 underwent microbiological testing, 210 ex vivo functional assessment, and 57 physicomechanical evaluation. No microorganism growth was detected in any evaluated device (100%, 95% CI 98.6-100). All but one overtube retained functionality (99.5%; 95% CI 97.4-99.9). Physicomechanical testing showed no clinically relevant differences compared with non-expired devices. Only six centres (24%) reported written protocols to minimise accumulation of expired accessories, and three (12%) had electronic tracking systems. Conclusions Expired endoscopic consumables are frequent and costly, with suboptimal inventory management within endoscopy units. In this multicentre bench and ex vivo study, analysed devices preserved sterility and functional performance beyond their labelled expiry dates. These findings challenge current shelf-life assumptions and support further regulatory evaluation, but should not be interpreted as evidence supporting routine clinical use of expired devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Five-year prospective microbiological evaluation of sterilised operating room supplies stored under event-related sterility maintenance conditions: a single-centre study with molecular characterisation of a fungal contaminantJournal of Hospital Infection · 2026
- Environmental, Economic, and Labour Burden of Time-related Sterility Maintenance and the Potential Reductions from Event-related Sterility Maintenance: A Single-Centre Gate-to-Gate Modelling Study.The Journal of hospital infection · 2026
- An 18-month pilot microbiological evaluation of sterilized operating room supplies stored under routine clinical conditions at a single hospital in JapanBMC Research Notes · 2026
- From cabinet to PlasmaBag: validating mobile endoscope storage with plasma technology.Microbiology spectrum · 2026
- Microbiological surveillance after automated reprocessing of gastrointestinal endoscopes in routine clinical practice: A prospective randomized microbiological surveillance study.PloS one · 2026
- Public disclosure and traceability of FDA-cleared remanufactured da Vinci instruments: a model-level documentary audit, 2022-2026.Journal of robotic surgery · 2026