Microbiological surveillance after automated reprocessing of gastrointestinal endoscopes in routine clinical practice: A prospective randomized microbiological surveillance study.
Masayuki Kato, Hideki Hayashi, Nobuharu Hironaka, Masafumi Chiba, Takahiro Abe, Nana Shimamoto, Mamoru Ito, Masakuni Kobayashi,
Department of Endoscopy, The Jikei University School of Medicine, Tokyo, Japan. · Kaigen Pharma Co., Ltd., Osaka, Japan.
Key points
From the paper's conclusion- We conducted an exploratory, single-center, prospective randomized surveillance study of 222 endoscope-use episodes involving patient-used forward-viewing upper and lower gastrointestinal endoscopes.
Abstract
Microbiological surveillance is used to monitor the quality of gastrointestinal endoscope reprocessing, but prospective data comparing automated endoscope reprocessor systems under routine workflow conditions remain limited. We conducted an exploratory, single-center, prospective randomized surveillance study of 222 endoscope-use episodes involving patient-used forward-viewing upper and lower gastrointestinal endoscopes. Endoscope-use episodes were allocated 1:1:1 to strongly acidic electrolyzed water, ozonated water, or peracetic acid using an Excel-generated random-number procedure (n = 74 each). Assigned reprocessing was performed after standardized bedside pre-cleaning and manual cleaning. Post-reprocessing samples were obtained from predefined external and internal sites. Surveillance success was defined as ≤20 colony-forming units per sample at all sampled sites and absence of predefined indicator organisms. Success was achieved in 67/74 (90.5%) strongly acidic electrolyzed-water episodes, 71/74 (95.9%) ozonated-water episodes, and 71/74 (95.9%) peracetic-acid episodes. Risk differences in surveillance success versus peracetic acid were -5.4 percentage points (95% confidence interval, -14.6 to 3.2) and 0.0 percentage points (95% confidence interval, -7.7 to 7.7), respectively. In exploratory post hoc non-inferiority analyses using a -5 percentage-point margin, non-inferiority was not demonstrated for either functional-water system. A -10 percentage-point margin supported ozonated water but not strongly acidic electrolyzed water. Post hoc scope-level sensitivity analyses were directionally consistent with the primary analysis and highlighted the influence of repeated failures in an individual physical endoscope. These findings should be interpreted as exploratory microbiological quality-assurance data, not as evidence of clinical infection risk or comparative clinical effectiveness. Study registration: University Hospital Medical Information Network Clinical Trials Registry (UMIN-CTR), UMIN000061627 (registered May 19, 2026; enrollment began November 8, 2024).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Three-hospital study of endoscope protein contamination and microbiome.American journal of infection control · 2026
- Development and internal validation of a risk-scoring model for gastrointestinal flexible endoscope reprocessing failure: A retrospective observational study.The Journal of international medical research · 2026
- Diagnostic and Clinical Performance of Single-Use Versus Reusable Gastrointestinal Endoscopes: A Systematic Review and Meta-Analyses.Journal of gastroenterology and hepatology · 2026
- A PBMC-based assay detects residual pro-inflammatory response in endoscopes after high-level disinfection undetected by ATP bioluminescence: a pilot study.The Journal of hospital infection · 2026
- Antibiotic resistance of bacteria isolated from microbiological control of gastrointestinal endoscopes.The Journal of hospital infection · 2026
- ESKAPE-Oriented Microbiological Quality Control of Reprocessed Mechanical Ventilators: Implications for Ventilator-Associated Pneumonia and Economic EvaluationMicrobiology Research · 2026