Response of the peeled off Gafchromic EBT2 film to 4 MeV alpha particles using transmission and reflection modes.
Hala I El-Naggar, M S Ghaith, M El Ghazaly, A Al-Sayed, Nermeen A Kelany, Essam E Abdel Fattah, Mostafa A M Abdeen, Amal Aboelmaaref,
Department of Physics, Faculty of Science, Benha University, Benha, 13518, Egypt. halaalnaggarr@gmail.com. · Department of Physics, Faculty of Science, Zagazig University, Zagazig, 44519, Egypt.
Key points
From the paper's conclusion- For the peeled-off Gafchromic EBT2 film, the spatial resolution in terms of full width at half maximum (FWHM) shows a slight dose independence on 4 MeV alpha particle doses for all color channels in transmission mode, but it is slightly correlated with the reflection mode.
Abstract
The dose-response and spatial resolution of the peeled-off Gafchromic EBT2 films irradiated with different doses of 4 MeV alpha particles were characterized using transmission-and-reflection scanning modes of a flatbed scanner. The Edge Spread Function (ESF) was produced by positioning a razor blade with a sharp edge directly on the film's active layer to minimize the penumbra. Subsequently, 4 MeV alpha particles were incident normally on the setup, producing two distinct regions on the film. The irradiated peeled-off Gafchromic EBT2 films with 4 MeV alpha particles were digitized using a flatbed scanner in both transmission and reflection modes under identical conditions. In reflection mode, the red and green channel pixel values exhibited a more rapid decay compared to those in transmission mode. The edge spread function (ESF) data was fitted using an error function, and which is then differentiated to obtain the line spread function (LSF). In transmission mode, the Weber contrast increased significantly with dose, compared to the reflection mode. For the peeled-off Gafchromic EBT2 film, the spatial resolution in terms of full width at half maximum (FWHM) shows a slight dose independence on 4 MeV alpha particle doses for all color channels in transmission mode, but it is slightly correlated with the reflection mode. In both transmission and reflection scanning modes, the green channel exhibits the highest spatial resolution. The spatial frequency was determined at the half-maximum of the modulation transfer function (MTF-50), with the green channel yielding the highest spatial frequency in both modes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of scanning position and orientation on the readout of Gafchromic film scanned with two different light-emitting flatbed scanners.Radiation and environmental biophysics · 2026
- Comparative analysis of the responses of EBT-XD and EBT4 films to ultra-high dose rate proton beamsRadiation Physics and Chemistry · 2025
- Explicit pixel-value correction of lateral response artifacts in radiochromic film dosimetry: Evaluation of single-, double-, and triple-channel methods.ArXiv · 2026
- Humidity effects in GafChromicTMEBT4 films for postal audits: characterization, reversibility and exploration of correction methods.Physics in medicine and biology · 2026
- Radiation-Sensitive Thin Film Dosimeter Based on Polyvinyl Alcohol (PVA)/Hafnium Dioxide (HfO2)/Silver Nitrate (AgNO3) Composite: Colorimetric Characterization and Dose-Response Analysis for Low-Dose Gamma-Ray Applications.Polymers · 2026
- The influence of hypoxia and hypercapnia on Gafchromic™ MD-V3 and HD-V2 film response during 60Co gamma irradiation.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026