Effect of scanning position and orientation on the readout of Gafchromic film scanned with two different light-emitting flatbed scanners.

Sheetal Sharma, Ranjit Singh, Divya Khosla, Shikha Goyal

Department of Radiotherapy and Oncology, Post Graduate Institute of Medical Education and Research, Chandigarh, 160012, India. · Department of Radiotherapy and Oncology, Post Graduate Institute of Medical Education and Research, Chandigarh, 160012, India. ranjit783@gmail.com.

Key points

From the paper's conclusion
  • The results demonstrate that scanner response is non-uniform across the scan area and is influenced by film orientation, position, irradiation dose, and scanner light source, with effects more pronounced at lower doses.

Abstract

The present study investigates the dose readout characteristics of Gafchromic EBT3 films over a dose range of 0.5-5 Gy using two flatbed scanners in-built with different light sources, Epson Expression 11000XL and 12000XL. Film readouts were acquired repeatedly at varying scan positions and film orientations to evaluate scanner-related non-uniformities. At 0.5 Gy, the maximum variation in optical density over the entire scan frame was 15.5%, 13%, and 29%, for the red, green, and blue channels, respectively, for the Epson 11000XL, compared with 9.5%, 7%, and 22.3%, for the Epson 12000XL. At 5 Gy, the corresponding maximum differences were 2.5%, 4.0%, and 4.1%, for the 11000XL, and 6.7%, 2.5%, and 10.1%, for the 12000XL. Changes in film orientation relative to the scan direction introduced additional variability. For the Epson 11000XL, the highest orientation-related differences at 0.5 Gy and 5 Gy were 15.3%, 2.1%, and 10.4%, and, 6.3%, 3.6%, and 1.3%, for the red, green, and blue channels, respectively. In the Epson 12000XL, the corresponding differences were higher at low dose, reaching 8.2%, 7.0%, and 26.8% at 0.5 Gy, and 5.5%, 8.5%, and 4.7% at 5 Gy. The results demonstrate that scanner response is non-uniform across the scan area and is influenced by film orientation, position, irradiation dose, and scanner light source, with effects more pronounced at lower doses. LED-based scanners exhibit marginally superior performance compared with fluorescent lamp-based systems. Consistent film orientation and position during scanning is therefore recommended. A standard operating flowchart for scanner calibration and quality assurance is proposed.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Effect of scanning position and orientation on the readout of Gafchromic film scanned with two different light-emitting flatbed scanners — Science Skim