Explicit pixel-value correction of lateral response artifacts in radiochromic film dosimetry: Evaluation of single-, double-, and triple-channel methods.
Lixiang Guo, Ken Kang-Hsin Wang
Key points
From the paper's conclusion- LRA should be corrected in the raw-PV domain before dose reconstruction. Full correction better preserved profile consistency and dose accuracy, particularly for EBT-XD. Because improved profile consistency did not always improve central-dose agreement, both should be evaluated separately in commissioning.
Abstract
BACKGROUND: Flatbed scanners introduce a lateral response artifact (LRA) perpendicular to lamp motion. Because LRA varies with position, pixel value (PV), color channel, and film model, it may violate the channel-common perturbation assumption of multichannel dosimetry. PURPOSE: To characterize PV-domain LRA in EBT4 and EBT-XD films, compare Lewis and Full corrections, and evaluate their effects on single-, double-, and triple-channel dosimetry. METHODS: Seven films were scanned at seven lateral positions using a cyclic-shift design. For each position and RGB channel, center-equivalent response was modeled as PVc = A + B * PV. Lewis correction used two films with linear interpolation of A and B; Full correction used all center-local pairs with PCHIP interpolation of PVc - PV. Performance was assessed by portrait-landscape profile consistency and dose accuracy. RESULTS: LRA was affine in PV, asymmetric about scanner center, and dependent on film model and channel. Both corrections reduced profile differences. Mean absolute profile differences across MU levels ranged from 1.4%-5.3% for EBT4 and 2.4%-6.8% for EBT-XD. The methods performed similarly for EBT4, while Full correction reduced discrepancies for several EBT-XD double- and triple-channel methods. Residual blue-channel errors limited blue, red-blue, and triple-channel performance. At 500 MU and above, red showed the best dose agreement; at 100 MU, both corrections increased deviations for several single-channel estimates. Full correction produced smaller deviations than Lewis in several EBT-XD conditions. CONCLUSIONS: LRA should be corrected in the raw-PV domain before dose reconstruction. Full correction better preserved profile consistency and dose accuracy, particularly for EBT-XD. Because improved profile consistency did not always improve central-dose agreement, both should be evaluated separately in commissioning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of scanning position and orientation on the readout of Gafchromic film scanned with two different light-emitting flatbed scanners.Radiation and environmental biophysics · 2026
- Response of the peeled off Gafchromic EBT2 film to 4 MeV alpha particles using transmission and reflection modes.Scientific reports · 2026
- Clinical in-vivo dosimetry for total skin electron therapy using gafchromic™ EBT4 film.Journal of applied clinical medical physics · 2026
- Microstructural changes in radiochromic film and dosimetry in magnetic fieldScientific Reports · 2026
- Comparative analysis of the responses of EBT-XD and EBT4 films to ultra-high dose rate proton beamsRadiation Physics and Chemistry · 2025
- Humidity effects in GafChromicTMEBT4 films for postal audits: characterization, reversibility and exploration of correction methods.Physics in medicine and biology · 2026