Maintaining setup reproducibility in patients unable to achieve bilateral arm abduction during breast radiotherapy: evaluation of a novel arm support device

Conor McGranaghan, Simon Goldsworthy, Robin Jhagra, Pauline Humphrey, Harley Stephens, Michael Spiro

Key points

From the paper's conclusion
  • Building on the Bi-Arm study framework, this prospective pilot evaluation assessed a Novel Arm Support Device (NASD) designed for unilateral arm abduction, examining its feasibility, geometric setup reproducibility, patient experience and operational impact.

Abstract

Introduction: Bilateral arm abduction remains the preferred standard for breast radiotherapy, but postoperative shoulder restrictions often necessitate alternative positioning. Building on the Bi-Arm study framework, this prospective pilot evaluation assessed a Novel Arm Support Device (NASD) designed for unilateral arm abduction, examining its feasibility, geometric setup reproducibility, patient experience and operational impact. Materials and methods: Between September 2025 and May 2026, 15 NASD patients and 15 matched historical controls were enrolled, each cohort contributing 105 imaged fractions. Daily image-guided radiotherapy (IGRT) quantified population systematic (Σ) and random (σ) error data. Patient-reported outcome measures (PROMs), fraction-level repeat imaging rates and therapeutic radiographer surveys were evaluated against the historical control cohort. Results: The NASD demonstrated high clinical feasibility. Translational errors and 3D vector displacement (7·72 ± 3·12 mm vs. 7·21 ± 2·85 mm; P = 0·334) broadly overlapped controls, while the NASD significantly stabilised systematic rotational pitch variations ( P = 0·046). PROMs yielded high patient acceptability, with median comfort scores of 4 (IQR: 4–5). At the fraction level, repeat imaging significantly decreased to 1·0% (1/105) for NASD versus 5·7% (6/105) for historical controls. Radiographers reported enhanced positioning consistency, with 87·5% reporting improved comfort and 14/16 demonstrating an overall preference for the NASD workflow. Conclusions: The NASD provides a practical, well-tolerated alternative that maintains geometric accuracy within standard tolerance frameworks for patients unable to tolerate bilateral positioning. Future research should prioritise a larger, multi-centre validation study with robust multivariable adjustments. Operationally, permanently dedicating an independent NASD assembly per active treatment room would optimise clinical workflows by eliminating inter-room scheduling delays.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers