Zeitschrift fur medizinische Physik

Can active bone conduction implant processors of two different systems be interchanged? A technical and clinical feasibility study.

Torsten Rahne, Dawid Brüning, Laura Fröhlich, Oliver Dziemba

University Medicine Halle, Department of Otorhinolaryngology, Head and Neck Surgery, Halle (Saale), Germany. Electronic address: torsten.rahne@uk-halle.de. · University Medicine Halle, Department of Otorhinolaryngology, Head and Neck Surgery, Halle (Saale), Germany.

Key points

From the paper's conclusion
  • The tested audio processors were technically capable of activating non-native implants, and short-term speech perception was possible in two supervised clinical assessments. Nevertheless, substantial differences in output characteristics indicate system-specific electro-mechanical optimization. However, differences in output characteristics and the lack of systematic electrical, thermal, and long-term safety evaluations mean that clinical interchangeability cannot be recommended. Further technical and clinical evaluation is required before cross-system use can be considered.

Abstract

OBJECTIVES: Active transcutaneous bone conduction implant systems are designed as closed hardware systems in which external audio processors operate exclusively with their corresponding implanted transducers. However, the electro-mechanical compatibility of components from different manufacturers has not previously been investigated. This study evaluated the technical feasibility, output performance, and preliminary clinical applicability of cross-system processor-implant combinations using the Bonebridge (MED-EL) and Sentio (Oticon Medical) systems. METHODS: A Bonebridge BCI 602 and a Sentio Ti implant were coupled to a skull simulator and implant phantoms. Native and non-native combinations of the Samba 2 BB and Sentio 1 Mini audio processors were evaluated under standardized reference test settings. Force output characteristics were measured using frequency response and maximum output recordings. In addition, two implanted patients underwent acute experimental fitting with the alternative processor system. Audiological assessment included aided sound-field thresholds, speech recognition in quiet and noise, categorical loudness scaling, and the Speech, Spatial and Qualities of Hearing Scale (SSQ12). RESULTS: Cross-system activation was technically possible in all investigated processor-implant combinations. Skull simulator measurements demonstrated substantial differences in output force levels between native and non-native pairings, with 4PTA force-output differences of up to 21.6 dB, although feedback-related interference affected measurements at high input levels. Phantom measurements confirmed audibility for all combinations. In both clinical assessments, successful fitting and open-set speech understanding were achieved with the non-native processor configuration. Speech recognition performance, aided thresholds, and subjective hearing outcomes were generally comparable between native and non-native configurations, with both participants reporting a preference for the Sentio 1 Mini processor in at least some outcome measures. CONCLUSIONS: The tested audio processors were technically capable of activating non-native implants, and short-term speech perception was possible in two supervised clinical assessments. Nevertheless, substantial differences in output characteristics indicate system-specific electro-mechanical optimization. However, differences in output characteristics and the lack of systematic electrical, thermal, and long-term safety evaluations mean that clinical interchangeability cannot be recommended. Further technical and clinical evaluation is required before cross-system use can be considered.

Read the paper