European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery

Clinical feasibility of a new fitting procedure for active bone conduction implants.

Miryam Calvino, Viktor Koci, Kamila Kondraciuk, Fiammetta M Korsch, Joachim Schmutzhard, Antonio Della Volpe, Hamidreza Mojallal, Javier Gavilán

Otolaryngology Department, IdiPAZ. La Paz University Hospital, Madrid, Spain. miryamcf@yahoo.com. · Clinic for Hearing, Speech and Voice Disorders, Medical University Hospital, Innsbruck, Austria.

Key points

From the paper's conclusion
  • 85.7% of test subjects preferred the new fitting formula over the standard fitting formula.

Abstract

PURPOSE: To compare, in users of Bonebridge (BB) system, the outcomes using two first fit algorithms, in an acute test setting with the patient's individually fine-tuned fitting based on standard fitting methods. METHODS: 22 individuals aged 5 years and older in the BB's indication range were included. Non-experienced patients were enrolled prospectively, and experienced BB users were included retrospectively. RESULTS: Mean sound field PTA4 thresholds were: 41 ± 7dB for standard, 32 ± 5dB for new and 32 ± 10 dB for own fittings (p < 0.001). Mean wor​d recognition scores at 45dB were: 22 ± 30% for standard, 71 ± 19% for new and 52 ± 40% for the patient's own fitting condition (p < 0.001). A ceiling effect was observed at 65dB (p = 0.36). The average results of the speech in noise tests were: -2.29 ± 1.97 dB SNR for standard, -4.11 ± 1.53 dB SNR for new and - 3.65 ± 1.77 dB SNR for the patient's own fitting conditions (p = 0.002). 85.7% of test subjects preferred the new fitting formula over the standard fitting formula. CONCLUSIONS: The new fitting formula demonstrated improved performance in both audiological and subjective measures in comparison to the standard. Further research is necessary to validate these findings in broader clinical settings.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Clinical feasibility of a new fitting procedure for active bone conduction implants — Science Skim