Comparison of sound processors for intraoperative monitoring of floating mass transducer coupling efficacy using auditory brainstem responses.
Oliver C Dziemba, Tobias Oberhoffner
University Medicine Greifswald, Department of Otorhinolaryngology, Head and Neck Surgery, Greifswald, Germany. Electronic address: https://www.unimedizin-greifswald.de/hno/. · MED-EL Deutschland GmbH, Starnberg, Germany.
Key points
From the paper's conclusion- The SAMBA 2 Hi is a suitable replacement for the AP 404. Due to the design of the SAMBA 2 Lo its output level and noise floor are lower compared to its Hi counter variant. That potentially offers options for testing individual programmed processors based on the degree of hearing loss. This idea should be investigated in further studies. The AcoustiAP currently does not allow reliable threshold-based assessment without correction.
Abstract
OBJECTIVE: The benefit of the Vibrant Soundbridge (VSB) active middle ear implant depends on the coupling quality between the floating mass transducer and the connection point in the middle ear. Intraoperative assessment via auditory brainstem responses has been proposed using either a programmed audio processor or a dedicated research tool (AcoustiAP). As the originally used processor (AP 404) is discontinued, this study compares the transmission characteristics of two successor devices (SAMBA 2 Hi, SAMBA 2 Lo) and the AcoustiAP against the AP 404. METHODS: The four devices (MED-EL, Innsbruck) were coupled to a 2-cc coupler of an Aurical HIT test box. CE-Chirp stimuli were presented at 0, 30 and 65 dB nHL; a 0-100 dB nHL input-output function was also recorded for the AcoustiAP. The sound processors were programmed using DSLv5 for a threshold of 25 dB HL. High-frequency averages (HFA) were calculated according to DIN EN 60118-7. RESULTS: Intrinsic noise HFA were 12.0, 22.5, 25.7 and 26.0 dB SPL (SAMBA 2 Lo, AP 404, AcoustiAP, SAMBA 2 Hi). At 30 dB nHL, the sound processoes produced comparable outputs (46.5-47.0 dB SPL), whereas the AcoustiAP remained within its noise floor. The SAMBA 2 Hi showed the most linear behaviour; the SAMBA 2 Lo showed pronounced compression. The AcoustiAP signal was embedded in noise up to about 35 dB nHL and saturated at 70.1 dB SPL. CONCLUSIONS: The SAMBA 2 Hi is a suitable replacement for the AP 404. Due to the design of the SAMBA 2 Lo its output level and noise floor are lower compared to its Hi counter variant. That potentially offers options for testing individual programmed processors based on the degree of hearing loss. This idea should be investigated in further studies. The AcoustiAP currently does not allow reliable threshold-based assessment without correction.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Can active bone conduction implant processors of two different systems be interchanged? A technical and clinical feasibility study.Zeitschrift fur medizinische Physik · 2026
- Twelve-Month Outcomes After Hearing Rehabilitation With an Active Middle Ear Implant and Latest Generation Vibroplasty Couplers in Adults And Children: A Prospective Multicenter Study.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Auditory rehabilitation in mixed hearing loss due to chronic otitis media sequelae: passive versus active transcutaneous bone conduction implant.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- A Prospective, Multicenter Clinical Investigation on the Safety and Performance of a New Active Transcutaneous Bone-Anchored Implant SystemOtology & Neurotology · 2025
- Assessing hearing and hearing loss in mice with OpenSABR (Open-source System for Auditory Brainstem Response).Journal of neurophysiology · 2026
- Safety and efficacy of current active transcutaneous bone conduction implants: a systematic review.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026