Comparison of sound processors for intraoperative monitoring of floating mass transducer coupling efficacy using auditory brainstem responses.

Oliver C Dziemba, Tobias Oberhoffner

University Medicine Greifswald, Department of Otorhinolaryngology, Head and Neck Surgery, Greifswald, Germany. Electronic address: https://www.unimedizin-greifswald.de/hno/. · MED-EL Deutschland GmbH, Starnberg, Germany.

Key points

From the paper's conclusion
  • The SAMBA 2 Hi is a suitable replacement for the AP 404. Due to the design of the SAMBA 2 Lo its output level and noise floor are lower compared to its Hi counter variant. That potentially offers options for testing individual programmed processors based on the degree of hearing loss. This idea should be investigated in further studies. The AcoustiAP currently does not allow reliable threshold-based assessment without correction.

Abstract

OBJECTIVE: The benefit of the Vibrant Soundbridge (VSB) active middle ear implant depends on the coupling quality between the floating mass transducer and the connection point in the middle ear. Intraoperative assessment via auditory brainstem responses has been proposed using either a programmed audio processor or a dedicated research tool (AcoustiAP). As the originally used processor (AP 404) is discontinued, this study compares the transmission characteristics of two successor devices (SAMBA 2 Hi, SAMBA 2 Lo) and the AcoustiAP against the AP 404. METHODS: The four devices (MED-EL, Innsbruck) were coupled to a 2-cc coupler of an Aurical HIT test box. CE-Chirp stimuli were presented at 0, 30 and 65 dB nHL; a 0-100 dB nHL input-output function was also recorded for the AcoustiAP. The sound processors were programmed using DSLv5 for a threshold of 25 dB HL. High-frequency averages (HFA) were calculated according to DIN EN 60118-7. RESULTS: Intrinsic noise HFA were 12.0, 22.5, 25.7 and 26.0 dB SPL (SAMBA 2 Lo, AP 404, AcoustiAP, SAMBA 2 Hi). At 30 dB nHL, the sound processoes produced comparable outputs (46.5-47.0 dB SPL), whereas the AcoustiAP remained within its noise floor. The SAMBA 2 Hi showed the most linear behaviour; the SAMBA 2 Lo showed pronounced compression. The AcoustiAP signal was embedded in noise up to about 35 dB nHL and saturated at 70.1 dB SPL. CONCLUSIONS: The SAMBA 2 Hi is a suitable replacement for the AP 404. Due to the design of the SAMBA 2 Lo its output level and noise floor are lower compared to its Hi counter variant. That potentially offers options for testing individual programmed processors based on the degree of hearing loss. This idea should be investigated in further studies. The AcoustiAP currently does not allow reliable threshold-based assessment without correction.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers