Auditory rehabilitation in mixed hearing loss due to chronic otitis media sequelae: passive versus active transcutaneous bone conduction implant.
Leonardo Elías Ordóñez Ordóñez, María Laura Rojas Serrano, José Agustín Caraballo Arias, Francisco Javier González Eslait, Juan Felipe Monroy Barreneche, José Eduardo Guzmán Durán, Rafael Jaramillo, Juan Carlos Izquierdo,
Department of Otolaryngology, Clínica Universitaria Colombia, Clínica Colsanitas SA, Keralty, Calle 23 #66‑46, Bogotá, Colombia. otoleor@gmail.com. · Hospital Militar Central, Universidad Militar Nueva Granada, Bogotá, Colombia.
Key points
From the paper's conclusion- Purpose To compare audiological outcomes and self‑perceived benefits in patients with sequelae of chronic otitis media (COM) and mixed hearing loss (MHL) undergoing rehabilitation with either a passive transcutaneous bone conduction implant (PT-BCI) or an active transcutaneous bone conduction implant (AT-BCI).
Abstract
PURPOSE: To compare audiological outcomes and self‑perceived benefits in patients with sequelae of chronic otitis media (COM) and mixed hearing loss (MHL) undergoing rehabilitation with either a passive transcutaneous bone conduction implant (PT-BCI) or an active transcutaneous bone conduction implant (AT-BCI). METHODS: A multicenter, prospective and observational analytical study in Colombia included patients receiving PT‑BCI or AT‑BCI. Main outcomes were pure‑tone average (PTA=0.5-4Khz), high‑frequency PTA (HF‑PTA=4-8 kHz), and Client Oriented Scale of Improvement (COSI). Comparative analyses used two‑way repeated‑measures ANOVA (complete‑case) followed by linear mixed‑effects models for repeated measures (LMMs‑RM) to address group imbalance and missing postoperative data, thereby optimizing statistical power. RESULTS: A total of 211 AT‑BCI and 50 PT‑BCI patients were included. Baseline characteristics were comparable, except for soft‑tissue thickness. Both devices achieved significant improvements in PTA and HF‑PTA at 3 and 12 months (p<0.010). LMMs‑RM confirming significant effects of device type (superior with AT‑BCI) and time for both outcomes: for PTA (Device type: F(1,321)=5.317, p=0.022; Time: F(2,321)=201.179, p<0.001) and for HF‑PTA (Device type: F(1,320)=5.665, p=0.018; Time: F(2, 320)=116.934, p<0.001). 82 patients (88.2%) reported that their performance with the device was "Better"/"Much better" on the COSI scale, no statistically significant differences observed between the AT‑BCI and PT‑BCI groups (p=0.283). CONCLUSIONS: The findings of the present study suggest that AT‑BCI provides a greater and clinically meaningful auditory gain for HF‑PTA in patients with MHL secondary to COM. In ATBCI group a stable audiological performance was achieved as early as the 3‑month follow‑up and is maintained over time.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Prospective, Multicenter Clinical Investigation on the Safety and Performance of a New Active Transcutaneous Bone-Anchored Implant SystemOtology & Neurotology · 2025
- Safety and efficacy of current active transcutaneous bone conduction implants: a systematic review.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- Twelve-Month Outcomes After Hearing Rehabilitation With an Active Middle Ear Implant and Latest Generation Vibroplasty Couplers in Adults And Children: A Prospective Multicenter Study.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Can active bone conduction implant processors of two different systems be interchanged? A technical and clinical feasibility study.Zeitschrift fur medizinische Physik · 2026
- Bone conduction implants in Australia over two decades.Cochlear implants international · 2026
- Clinical feasibility of a new fitting procedure for active bone conduction implants.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026