Endoscopic suturing and flange modification of a biliary lumen-apposing metal stent for reduction of solid-food biliary reflux.
Kambiz Kadkhodayan
Center for Interventional Endoscopy, AdventHealth, Orlando, Florida, USA.
Key points
From the paper's conclusion- This article illustrates a technical modification of a biliary LAMS that may reduce solid luminal reflux. Further evaluation is required to define durability, safety, and broader applicability.
Abstract
BACKGROUND AND AIMS: EUS-guided biliary drainage (EUS-BD) using lumen-apposing metal stents (LAMSs) is an established approach for biliary access in patients with inaccessible ampullas or surgically altered anatomy. A recognized limitation of EUS-BD is stent dysfunction related to luminal solid debris, which may be particularly problematic in patients with impaired gastric emptying. We describe a novel technical modification involving partial closure of the proximal flange of a biliary LAMS using endoscopic suturing to reduce the retrograde passage of solid luminal contents while preserving antegrade bile flow. METHODS: A 60-year-old female with a history of postcholecystectomy bile duct injury requiring hepaticoduodenostomy presented with recurrent cholangitis over several years. Prior interventions included repeated ERCPs and placement of a biliary LAMS with coaxial plastic stents, yet solid-food debris was encountered within the biliary tree. During a repeat intervention, an 8- × 6-mm LAMS was placed, followed by partial closure of the proximal flange using endoscopic suturing. RESULTS: Liver chemistry results normalized, and functional imaging demonstrated unrestricted biliary drainage without evidence of retrograde solid-food entry into the bile ducts. CONCLUSIONS: This article illustrates a technical modification of a biliary LAMS that may reduce solid luminal reflux. Further evaluation is required to define durability, safety, and broader applicability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Advanced endoscopic rescue after failed endoscopic ultrasound-guided choledochoduodenostomy.VideoGIE : an official video journal of the American Society for Gastrointestinal Endoscopy · 2026
- Failure of release of the Hot AXIOS distal flange: "handle springing" as a rescue techniqueEndoscopy · 2026
- Electrocautery-Enhanced Lumen-Apposing Metal Stents Versus Modified Tubular Metal Stents for EUS-Guided Gallbladder Drainage: A Multicenter Study.Digestive endoscopy : official journal of the Japan Gastroenterological Endoscopy Society · 2026
- Safety and Outcomes of EUS-Guided Choledochoduodenostomy in Patients with a Common Bile Duct Diameter ≤15 mm.Gastrointestinal endoscopy · 2026
- Through-the-scope fully covered self-expandable metal stent with biflanged ends for ampullary access in severe duodenal obstruction.VideoGIE : an official video journal of the American Society for Gastrointestinal Endoscopy · 2026
- Externally anchored fully covered metal stents with plastic stent versus uncovered metal stents for unresectable distal malignant biliary obstruction: a multicenter randomized trial.Endoscopy · 2026