Externally anchored fully covered metal stents with plastic stent versus uncovered metal stents for unresectable distal malignant biliary obstruction: a multicenter randomized trial.
Jin Ho Choi, Junyeol Kim, Woo Hyun Paik, In Rae Cho, Ji Kon Ryu, Sung Yong Han, Seong-Hun Kim, Eaum Seok Lee,
Seoul National University Hospital, Seoul National University College of Medicine, Department of Internal Medicine and Liver Research Institute, Korea (the Republic of), Seoul. · Pusan National University Hospital, Division of Gastroenterology, Department of Internal Medicine, Biomedical Research Institute, Korea (the Republic of), Busan.
Key points
From the paper's conclusion- FCSEMS-EPS significantly reduced stent dysfunction compared with USEMS without increasing adverse events, supporting its effectiveness and safety as long-term biliary drainage in unresectable MBO.
Abstract
BACKGROUND: Self-expandable metal stents (SEMS) are commonly used to treat malignant biliary obstruction (MBO). External plastic stent anchoring for fully covered SEMS (FCSEMS-EPS) has shown potential to improve stent patency; however, randomized evidence comparing this strategy with uncovered SEMS (USEMS) is limited. METHODS: This multicenter trial was conducted at nine tertiary referral centers in South Korea. Patients were enrolled from October 2021 to February 2024 and followed for 12 months after stent placement. Those with unresectable distal MBO were randomly assigned (1:1) to receive either FCSEMS-EPS or USEMS. The primary outcome was biliary stent dysfunction. Time-to-event analyses were performed using time to stent dysfunction within a competing risk framework. RESULTS: 113 patients were enrolled, of whom 99 were included in the final analysis (FCSEMS-EPS 48; USEMS 51). Stent dysfunction occurred less frequently in the FCSEMS-EPS group compared with the USEMS group (27.1% vs. 49.0%; P = 0.04). The cumulative incidence of stent dysfunction at 6 months was lower in the FCSEMS-EPS group (21.1% vs. 29.8%; Gray's test, P = 0.03). In multivariable competing risk analysis, FCSEMS-EPS was independently associated with a lower risk of stent dysfunction (subdistribution hazard ratio 0.44, 95%CI 0.21-0.90; P = 0.03). Procedure-related adverse events and overall survival did not differ between groups. CONCLUSION: FCSEMS-EPS significantly reduced stent dysfunction compared with USEMS without increasing adverse events, supporting its effectiveness and safety as long-term biliary drainage in unresectable MBO.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of Uncovered and Covered Biliary SEMS for the Salvage of Obstructed SEMS: A Multicenter Retrospective Propensity Score-Weighted Study.Journal of gastroenterology and hepatology · 2026
- Multi-hole Versus Fully Covered Self-expandable Metal Stents for Distal Biliary Obstruction in Unresectable Pancreatic Cancer.Pancreas · 2026
- Types of Self-Expandable Metal Stents for Palliative Drainage of Unresectable Extrahepatic Malignant Biliary Obstruction: A Network Meta-Analysis of Randomized Controlled Trials.Endoscopy · 2026
- Comparative Effectiveness of Two Novel Anti-Migration Strategies in Fully Covered Self-Expandable Metal Stents for Biliary Strictures: A Multicenter Propensity-Matched Study.Gastrointestinal endoscopy · 2026
- Outcomes of a novel biliary hybrid metal stent composed of an inner-covered and an outer-uncovered stent for unresectable distal malignant biliary obstruction (with video).Hepatobiliary & pancreatic diseases international : HBPD INT · 2026
- Predictors of dysfunction in fully versus partially covered self-expanding metal stents for malignant esophageal strictures: a 10-year multicenter study.BMC gastroenterology · 2026