Evaluation of an anaesthetic conserving device in semi-closed rebreathing systems using an artificial lung with isoflurane and sevoflurane: an in vitro pilot study.
Jana Flyps, Maarten Haspeslagh, Stefaan Bouchez, Stijn Schauvliege
Department of Large Animal Surgery, Anaesthesia and Orthopaedics, Faculty of Veterinary Medicine, University of Ghent, Merelbeke-Melle, Belgium. Electronic address: jana.flyps@ugent.be. · Department of Large Animal Surgery, Anaesthesia and Orthopaedics, Faculty of Veterinary Medicine, University of Ghent, Merelbeke-Melle, Belgium.
Key points
From the paper's conclusion- At low VTs (≤500 mL), the ACD adsorbed efficiently, with no (isoflurane) to minimal (≤20%) (sevoflurane) loss detected. As VT increased (≥800 mL), breakthrough became greater, ranging from 20% to 28% and 40% to 53% for isoflurane and sevoflurane, respectively. The highest recirculation-adjusted breakthrough occurred at 6000 mL with 10 cmH2O PEEP for isoflurane (28%) and at 800 mL with 10 cmH2O PEEP for sevoflurane (53%). Sevoflurane showed higher breakthrough than isoflurane for all settings.
Abstract
OBJECTIVE: To evaluate the performance of an anaesthetic conserving device (ACD), in semi-closed rebreathing systems. STUDY DESIGN: In vitro pilot study. ANIMALS: None. METHODS: First, an ACD was positioned between an artificial lung and the Y-piece of a human anaesthesia machine with a ventilator, delivering tidal volumes (VTs) of 350, 500, 800, and 1200 mL. Second, using a three-dimensionally printed connector, four ACDs were placed in parallel between the artificial lung and the Y-piece of a large animal anaesthesia machine with a ventilator, delivering VTs of 4000 and 6000 mL. Each VT, delivered during volume-controlled ventilation, was tested using isoflurane and sevoflurane and using 0 and 10 cmH2O of positive end-expiratory pressure (PEEP). Anaesthetic agents were delivered via continuous infusion, aiming for end-tidal concentrations of 1.0% (isoflurane) and 2.0% (sevoflurane). Inspiratory and expiratory vapour concentrations were monitored continuously at the ACD sampling port and at the Y-piece over a 30-minute period. The anaesthetic agent breakthrough (the proportion of exhaled anaesthetic vapour not captured by the ACD) was calculated for each setting and adjusted for the effect of recirculation (recirculation-adjusted breakthrough). RESULTS: At low VTs (≤500 mL), the ACD adsorbed efficiently, with no (isoflurane) to minimal (≤20%) (sevoflurane) loss detected. As VT increased (≥800 mL), breakthrough became greater, ranging from 20% to 28% and 40% to 53% for isoflurane and sevoflurane, respectively. The highest recirculation-adjusted breakthrough occurred at 6000 mL with 10 cmH2O PEEP for isoflurane (28%) and at 800 mL with 10 cmH2O PEEP for sevoflurane (53%). Sevoflurane showed higher breakthrough than isoflurane for all settings. CONCLUSIONS AND CLINICAL RELEVANCE: This is the first study to evaluate an ACD in semi-closed rebreathing systems. The results show that although the ACD effectively captures volatile anaesthetics at low VTs, its breakthrough increased at higher VTs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Combining an anaesthetic conserving device with a circle breathing system to minimise isoflurane waste: an in vitro study.Veterinary anaesthesia and analgesia · 2026
- Utilization and Estimation of Environmental Impact of End-Tidal Control in a Multisite Academic Health System: A Retrospective Observational Cross-Sectional Study.A&A practice · 2026
- Comparison of bag-valve-device ventilation and volume-controlled ventilation with laryngeal tube evo in intra-arrest ventilation-a prospective randomized study in human body donors.Journal of anesthesia, analgesia and critical care · 2026
- Washout characteristics of sevoflurane and desflurane in surgical patients: a prospective observational study using gas chromatography - ion mobility spectrometry.BMC anesthesiology · 2026
- Airway Pressure Release Ventilation with a Time-Controlled Adaptive Ventilation Strategy (APRV-TCAV) Reverses Endotracheal Secretion Movement Compared with Volume-Control Ventilation: A Bench Study.Journal of clinical medicine · 2026
- Automated Blood Conditioning for Artificial Lung Evaluation: A Feasibility Study.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026