Airway Pressure Release Ventilation with a Time-Controlled Adaptive Ventilation Strategy (APRV-TCAV) Reverses Endotracheal Secretion Movement Compared with Volume-Control Ventilation: A Bench Study.
Ben T Daxon, William M LeTourneau, Brendan T Wanta, Mariah L Bennett, Abier M Dawood, Rylee N Stewart, Pia P McEleney
Department of Anesthesiology and Perioperative Medicine, Mayo Clinic, 200 First Street SW, Rochester, MN 55905, USA. · Department of Respiratory Care, Mayo Clinic, 200 First Street SW, Rochester, MN 55905, USA.
Key points
From the paper's conclusion- Airway Pressure Release Ventilation following a Time-Controlled Adaptive Ventilation protocol (APRV-TCAV) has been anecdotally observed to improve secretion clearance, but the mechanism is not well established.
Abstract
Background/Objectives: Retained airway secretions are a common complication of mechanical ventilation and a major contributor to ventilator-associated pneumonia. Airway Pressure Release Ventilation following a Time-Controlled Adaptive Ventilation protocol (APRV-TCAV) has been anecdotally observed to improve secretion clearance, but the mechanism is not well established. Because TCAV sets the release time (TLow) from each patient's own expiratory flow decay, we hypothesized that the resulting expiratory-dominant flow profile would generate a net antegrade force on secretions. This bench study compared secretion movement between APRV-TCAV and conventional Volume-Control Continuous Mandatory Ventilation (VC-CMV), with and without an in-line oscillatory device. Methods: A critical care ventilator was connected to a 3 L test lung via a 7 mm endotracheal tube (ETT). Simulated mucus (1% guar gum) was instilled into the ETT, and secretion movement was measured under VC-CMV (positive end-expiratory pressure (PEEP) 5, 10, 15 cmH2O) and APRV-TCAV (PHigh 20, 25, 30 cmH2O; PLow 0 cmH2O). Each condition was repeated with an in-line oscillatory device. Results: All VC-CMV settings produced retrograde secretion movement toward the test lung; all APRV-TCAV settings produced antegrade movement. The largest antegrade movement was observed at the widest PHigh-to-PLow differential. Adding the oscillatory device raised peak expiratory flow above peak inspiratory flow in every condition, yet abolished antegrade movement under APRV-TCAV. Conclusions: In this bench study, APRV-TCAV reversed the direction of secretion movement within the ETT, with larger antegrade movement at wider PHigh-to-PLow differentials. This suggests a potential mechanism by which the ventilator flow profile itself may influence secretion movement. The directional effect depended on the intact release-phase waveform and was abolished by in-line oscillatory therapy despite preserved peak flow asymmetry.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Computational Analysis of Unsteady Airway Flow Physics During Airway Pressure Release Ventilation.Journal of biomechanical engineering · 2026
- Expert recommendations for setting and adjusting airway pressure release ventilation based on clinical experience and basic science evidenceFrontiers in Medicine · 2026
- Reverse Triggering With and Without Breath Stacking During Pressure-Controlled Ventilation: Exposure Burden and Associations With Ventilation OutcomesRespiratory Care · 2026
- The Missing Dimension in Lung-Protective Ventilation: Physiologically Directed Inspiratory and Expiratory Time.Journal of clinical medicine · 2026
- Continuous versus intermittent endotracheal tube cuff pressure control for the prevention of ventilator-associated pneumonia: A focused systematic review and exploratory meta-analysis.Saudi journal of anaesthesia · 2026
- Flow-controlled ventilation versus pressure-controlled ventilation in moderate to severe ARDS patients: a randomized crossover physiological studyIntensive Care Medicine Experimental · 2025