Automated Blood Conditioning for Artificial Lung Evaluation: A Feasibility Study.
Louis William Beardell, Isabella Haladjian, Alisha Malik, Jingrui Liu, Vinisha Somaya, Alvaro Rojas-Peña, Joseph A Potkay
From the Department of Surgery, Robert H. Bartlett Extracorporeal Life Support Laboratory, University of Michigan, Ann Arbor, Michigan.
Key points
From the paper's conclusion- Automated blood conditioning is a closed-loop perfusion circuit, control algorithm, and protocol that rapidly conditions and continuously maintains blood conditions for artificial lung and ECLS system evaluation, permitting efficient testing of ECLS devices with minimal blood volume, and showing potential for long-term use.
Abstract
Artificial organs and Extracorporeal Life Support (ECLS) systems require extensive in vitro evaluation before in vivo preclinical and human trials are conducted. Current methodologies require large volumes of blood, extensive manual labor, and lengthy conditioning periods. To remedy these shortcomings, an "Automated Blood Conditioning (ABC) platform" was developed. Automated blood conditioning is a closed-loop perfusion circuit, control algorithm, and protocol that rapidly conditions and continuously maintains blood conditions for artificial lung and ECLS system evaluation, permitting efficient testing of ECLS devices with minimal blood volume, and showing potential for long-term use. The platform includes a hemoconcentrator, Terumo CDI 500 blood parameter monitor (CDI 500), custom gas mixer, and computer-based controller. A proportional-integral-derivative (PID) algorithm automatically adjusts sweep gas concentrations to maintain target inlet blood conditions despite disturbances from the test device. From arterial conditions, oxygen saturation (sO2%) settles at the venous target of 65% in approximately 5 minutes, and partial pressure of carbon dioxide (pCO2) at 45 mm Hg in approximately 6 minutes. In this feasibility study, four independent runs on separate blood batches (48.5 hours) held sO2% within ±5% of target for 99.2% of steady-state samples (error 1.2%); the primary evaluation run maintained physiologic conditions for its first 10 closed-loop hours, and device flows up to 0.6 L/min using 1.1 L of blood.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pulse Physiology Engine to simulate transient oxygen changes and closed-loop, autonomous pump control during venovenous extracorporeal membrane oxygenation.Research square · 2026
- Automatic control of arterial blood pressure: computer, benchtop, and animal models.Frontiers in medical technology · 2026
- Challenges and opportunities in ECMO monitoring: A mixed-methods appraisal.The journal of extra-corporeal technology · 2026
- Reperfused human umbilical cords: hydrodynamic and hemolytic analysis of the ex vivo model for umbilical cord ECMO cannulation.Frontiers in bioengineering and biotechnology · 2026
- Oxygenator deterioration in a standardized in vitro setup: pressure drop, CT-based clot burden, and hemolysisCurrent Directions in Biomedical Engineering · 2026
- Automated Control of Oxygen Therapy: A State-of-the-Art Review of Closed-Loop Devices, with a Rule-Based, Network-Independent System (EMILY) as a Working Exemplar.Seminars in respiratory and critical care medicine · 2026