Automated Blood Conditioning for Artificial Lung Evaluation: A Feasibility Study.

Louis William Beardell, Isabella Haladjian, Alisha Malik, Jingrui Liu, Vinisha Somaya, Alvaro Rojas-Peña, Joseph A Potkay

From the Department of Surgery, Robert H. Bartlett Extracorporeal Life Support Laboratory, University of Michigan, Ann Arbor, Michigan.

Key points

From the paper's conclusion
  • Automated blood conditioning is a closed-loop perfusion circuit, control algorithm, and protocol that rapidly conditions and continuously maintains blood conditions for artificial lung and ECLS system evaluation, permitting efficient testing of ECLS devices with minimal blood volume, and showing potential for long-term use.

Abstract

Artificial organs and Extracorporeal Life Support (ECLS) systems require extensive in vitro evaluation before in vivo preclinical and human trials are conducted. Current methodologies require large volumes of blood, extensive manual labor, and lengthy conditioning periods. To remedy these shortcomings, an "Automated Blood Conditioning (ABC) platform" was developed. Automated blood conditioning is a closed-loop perfusion circuit, control algorithm, and protocol that rapidly conditions and continuously maintains blood conditions for artificial lung and ECLS system evaluation, permitting efficient testing of ECLS devices with minimal blood volume, and showing potential for long-term use. The platform includes a hemoconcentrator, Terumo CDI 500 blood parameter monitor (CDI 500), custom gas mixer, and computer-based controller. A proportional-integral-derivative (PID) algorithm automatically adjusts sweep gas concentrations to maintain target inlet blood conditions despite disturbances from the test device. From arterial conditions, oxygen saturation (sO2%) settles at the venous target of 65% in approximately 5 minutes, and partial pressure of carbon dioxide (pCO2) at 45 mm Hg in approximately 6 minutes. In this feasibility study, four independent runs on separate blood batches (48.5 hours) held sO2% within ±5% of target for 99.2% of steady-state samples (error 1.2%); the primary evaluation run maintained physiologic conditions for its first 10 closed-loop hours, and device flows up to 0.6 L/min using 1.1 L of blood.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers