Performance Characteristics for Flow Cytometric Enumeration of Peripheral Blood Lymphocyte Subsets under Harmonized Laboratory Conditions: A Multicenter Study.

Lin Xiao, Yuan Wang, Yi Li, Feng Wang, Liming Cheng, Chaohui Duan, Min Hu, Mianyang Li,

Department of Clinical Laboratory, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China. · Department of Laboratory Medicine/Clinical Laboratory Medicine Research Center, West China Hospital, Sichuan University; Sichuan Clinical Research Center for Laboratory Medicine, Chengdu, Sichuan, China.

Key points

From the paper's conclusion
  • This study establishes a comprehensive performance threshold and dual-criteria evaluation system covering precision, comparability, and stability for EPBLS, based on a harmonized multicenter protocol using stratified routine clinical samples, providing a practical reference for laboratory validation and quality specification development.

Abstract

BACKGROUND: Flow cytometric enumeration of peripheral blood lymphocyte subsets (EPBLS) is essential for disease management, yet standardized validation criteria are lacking. We defined key performance characteristics under harmonized conditions to support criteria development. METHODS: 205 whole blood samples and 54 quality control samples at two levels from nine clinical laboratories across five regions in China were used to measure absolute counts (AC) and relative counts (RC) of lymphocyte subsets using flow cytometric single-platform. Performance characteristics were evaluated based on 2,865 measurements and an integrated statistical approach combining the concordance correlation coefficient, Passing-Bablok regression, and Bland-Altman plots. RESULTS: Performance characteristics for precision, comparability, and sample stability were established to cover the major clinically relevant lymphocyte subset distributions (RC: 1.32-97.53%; AC: 3.37-6,846.98 cells/μL). Intra-assay and inter-assay coefficients of variation were below 8% and 6% for RC, below 10% and 9% for AC, respectively. A rigorous dual-criteria framework (comparability criteria: consistency and correlation; deviation criteria: absolute maximum bias threshold of <20%), supported by a triple-statistical approach, confirmed agreement across reagent batches, laboratories, and storage time points, with comprehensive bias profiles further reported. Whole blood remained stable for 24 hours post-collection (RC bias <14%; AC bias ≤15%), and processed samples for 48 hours post-staining (RC bias <13%; AC bias <19%). CONCLUSIONS: This study establishes a comprehensive performance threshold and dual-criteria evaluation system covering precision, comparability, and stability for EPBLS, based on a harmonized multicenter protocol using stratified routine clinical samples, providing a practical reference for laboratory validation and quality specification development.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers