Analytical Performance of the Sysmex XF ‐1600 Flow Cytometer in Routine Clinical Practice
Meritxell Deulofeu Figueras, Anna Marull Arnall, Orlando Jiménez Romero, Natàlia Lloveras Guelque, David Cruz García, Leire Saiz Sierra, Maite Serrando Querol
Key points
From the paper's conclusion- This study aimed to evaluate the analytical performance of the novel Sysmex XF‐1600 flow cytometer for lymphocyte subpopulation quantification in peripheral blood, including precision, carryover, and agreement with established laboratory methods.
Abstract
Introduction Flow cytometry systems are increasingly used for the identification and quantification of lymphocyte subpopulations, supporting the diagnosis and monitoring of hematological and immunological conditions. This study aimed to evaluate the analytical performance of the novel Sysmex XF‐1600 flow cytometer for lymphocyte subpopulation quantification in peripheral blood, including precision, carryover, and agreement with established laboratory methods. Methods Peripheral blood samples from routine clinical workload were analysed in the hematology laboratory. Analytical performance was assessed by evaluating precision (repeatability and reproducibility) and carryover. Absolute lymphocyte and leukocyte counts obtained using the XF‐1600 automated sample preparation and volumetric counting system were compared with the Sysmex XN‐series hematology analyser, while lymphocyte subset quantification was compared with the Navios flow cytometer (Beckman Coulter). Results The XF‐1600 showed coefficients of variation below 6% for major lymphocyte populations, including CD3+, CD4+, CD8+, CD19+, and CD16+/CD56+ cells. No carryover was detected. Comparison with the Sysmex XN analyser showed strong correlation for leukocyte and lymphocyte counts (Spearman's ρ = 0.986 and 0.993, respectively). Comparison with the Navios flow cytometer demonstrated high correlation across lymphocyte subpopulations (Spearman's ρ > 0.9 for most parameters). Bland–Altman and Passing–Bablok analyses showed good agreement between methods, although some parameters were not fully interchangeable. Conclusion The Sysmex XF‐1600 flow cytometer demonstrated good analytical performance for lymphocyte subset quantification, with high precision and strong agreement with routinely used laboratory methods. The integration of automated sample preparation and volumetric counting may support standardization of immunophenotyping workflows in clinical laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- FlowCytoMonitoR: An R-Shiny Program for Advanced Monitoring Flow Cytometer Performance.Cytometry. Part A : the journal of the International Society for Analytical Cytology · 2026
- Flow Cytometric Detection of Bone Marrow Cell Populations Phenotypically Compatible with Neuroblastoma: An Exploratory Comparison with Conventional Methods.Cancers · 2026
- Performance Characteristics for Flow Cytometric Enumeration of Peripheral Blood Lymphocyte Subsets under Harmonized Laboratory Conditions: A Multicenter Study.Translational research : the journal of laboratory and clinical medicine · 2026
- Case Study of a Cross-Platform Evaluation of an Optimized Multicolor Flow Cytometry Immunophenotyping Assay for Human Peripheral Whole Blood.Cytometry. Part A : the journal of the International Society for Analytical Cytology · 2026
- Utility of multiparametric flow cytometry in the diagnosis and staging of Classic Hodgkin lymphoma-Experience in real-world practice.Cytometry. Part B, Clinical cytometry · 2026
- Effects of different blood storage conditions on leukocytes cell population data using the Sysmex XN hematological analyzerBiochemia Medica · 2026