Case Study of a Cross-Platform Evaluation of an Optimized Multicolor Flow Cytometry Immunophenotyping Assay for Human Peripheral Whole Blood.
Lia Mogge, Udo Gaipl, Stefanie Corradini, Benjamin Frey, Anna-Jasmina Donaubauer
Translational Radiobiology, Department of Radiation Oncology, Uniklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany. · Department of Radiation Oncology, Uniklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
Key points
From the paper's conclusion- It is concluded that calculated absolute values cannot be compared between the two devices, even though the immunophenotyping assay can be successfully measured on the new DxFlex flow cytometer.
Abstract
Flow cytometry is a commonly used single-cell analysis technique that can be applied for the immunophenotyping of human blood samples. Nowadays, increasing numbers of clinical trials include translational research programs utilizing flow-based immune monitoring assays to assess patients' longitudinal immune status. Over time, flow cytometers require regular calibration, quality controls, maintenance, and need to be replaced by new devices after certain years of operation. When flow-based assays are utilized in clinical studies, it is important to transfer and re-establish them thoroughly on new devices to ensure stable measurements. This project describes a detailed and optimized immunophenotyping assay and its transfer from the Gallios flow cytometer to the DxFlex (both from Beckman Coulter). According to the manufacturer, the DxFlex is intended as a successor to the Gallios and features similar lasers and filters. This enables all antibodies and assayed populations to be represented on the new device. However, after calculating the concentrations of cell populations in the blood, deviations were observed between the two devices. It is concluded that calculated absolute values cannot be compared between the two devices, even though the immunophenotyping assay can be successfully measured on the new DxFlex flow cytometer. Switching this specific IPT assay is not feasible in ongoing clinical trials. When transferring a complex assay, parallel comparative measurements are required to ensure reliable data acquisition, population detection, and result comparability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Analytical Performance of the Sysmex XF ‐1600 Flow Cytometer in Routine Clinical PracticeInternational Journal of Laboratory Hematology · 2026
- A Flow Cytometry-Based Method for Profiling Immune Cells in Human Blood and Tumors.Journal of visualized experiments : JoVE · 2026
- A standardized single-tube 17-color spectral flow cytometry workflow for integrated immunophenotyping of human PBMCs and mixed co-culture systemsFrontiers in Immunology · 2026
- Applications of flow cytometry in equine medicine: current clinical and research perspectives.Frontiers in veterinary science · 2026
- FlowCytoMonitoR: An R-Shiny Program for Advanced Monitoring Flow Cytometer Performance.Cytometry. Part A : the journal of the International Society for Analytical Cytology · 2026
- Calibration of Flow Cytometers Enables Reproducible Measurements of Extracellular Vesicle Concentrations and Reference Range EstablishmentJournal of Extracellular Vesicles · 2025