A wearable sensor integrating a MXene-PLL-GOx-modified electrode and a Janus nanofibrous membrane for electrochemical monitoring of sweat glucose.
Qiujin Li, Xiaoxue Zhu, Jianing Ren, Jingran Su, Jixian Gong, Xiuming Liu, Kuzmichev Victor
School of Textile Science and Engineering, Tiangong University, Tianjin, 300387, China; Key Laboratory of Advanced Textile Composites, Ministry of Education, Tiangong University, Tianjin, 300387, China; National Innovation Center of Advanced Dyeing & Finishing Technology, Tai'an, Shandong, 271000, China; Shaoxing Keqiao Institute of Tiangong University, Shaoxing, Zhejiang, 312030, China. Electronic address: maldini@mail.nankai.edu.cn. · School of Textile Science and Engineering, Tiangong University, Tianjin, 300387, China; Key Laboratory of Advanced Textile Composites, Ministry of Education, Tiangong University, Tianjin, 300387, China; Shaoxing Keqiao Institute of Tiangong University, Shaoxing, Zhejiang, 312030, China.
Key points
From the paper's conclusion- The modified SPCE was integrated with the Janus nanofibrous membrane to fabricate a wearable sweat sensor, with the sensitivities of 16.70 μA mM-1 cm-2 for 0∼300 μM glucose levels, with a detection limit of 24.35 μM.
Abstract
MXene is a two-dimensional transition metal carbide with large specific surface area, good conductivity, and outstanding biocompatibility, which can be integrated with electrochemical devices, significantly enhancing the sensing performance. Herein, we present a wearable electrochemical sensor based on MXene composites modified electrode and Janus nanofibrous membranes for active sweat collection and precise detection of sweat glucose. Poly-l-lysine (PLL) is highly positively charged because of the -NH3+ groups, which acts as the crosslinker to bond MXene and glucose oxidase (GOx, containing -COO- groups) both negatively charged, forming a functional composite with high catalytic activity for glucose sensing. In this way, GOx was efficiently immobilized on the MXene nanosheets and then deposited on the screen-printed carbon electrode (SPCE). The intermediate H2O2 from enzymatic glucose oxidation was catalytically decomposed by MXene into H2O and O2. The recycled O2 further accelerated the enzymatic reaction on the SPCE. PU/PVA Janus nanofibrous membrane was prepared through electrospinning for active and effective sweat collection due to the unidirectional moisture conduction from its hydrophobic/hydrophilic side. The modified SPCE was integrated with the Janus nanofibrous membrane to fabricate a wearable sweat sensor, with the sensitivities of 16.70 μA mM-1 cm-2 for 0∼300 μM glucose levels, with a detection limit of 24.35 μM. The sensor also showed good stability, selectivity, and reproducibility, providing a new strategy for non-invasive sweat glucose monitoring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Interface-Programmed Growth of High-Aspect-Ratio 1D Ni-MOF Nanofibers on MXene for Wearable Enzyme-Free Glucose Sensing.ACS applied materials & interfaces · 2026
- Glucose Sensor Using Fe3O4 Functionalized MXene Nanosheets as a Promising Sensing Platform: Exploring the Potential of Electrochemical Detection of GlucoseChemosensors · 2026
- Investigation of a surface decorated Ag-MnO2@MGCN nanointerface towards membrane-free electrochemical Na+ ion sensing in sweat.RSC advances · 2026
- Asymmetric Heterostructured Breathable Janus Nanofibrous Membranes for Deep Learning-Assisted Wearable Sensing.Small (Weinheim an der Bergstrasse, Germany) · 2026
- MXene‐Enabled Wearable Biosensors: A Design Framework for Autonomous BiosensingChemElectroChem · 2026
- Advances in Wearable Technology: MXene-Based Multifunctional and Biomedical Smart TextilesACS Omega · 2025