Multi-center comparison of relative stopping power estimation of an ex vivo phantom using clinical X-ray CT scanners and a proton CT prototype.
Elena Fogazzi, Esther Bär, Mara Bruzzi, Mario Ciocca, Carlo Civinini, Paolo Farace, Francesco Fracchiolla, Callum Gillies,
Physics Department, University of Trento, Trento, Italy; Trento Institute for Fundamental Physics and Applications (TIFPA), Italian National Institute of Nuclear Physics (INFN), Trento, Italy. Electronic address: elena.fogazzi@curie.fr. · University College London Hospitals NHS Foundation Trust, London, UK.
Key points
From the paper's conclusion- This work presents the first inter-center comparison of xCT-based RSP estimation using an ex vivo phantom and pCT measurements as reference. The observed variability highlights the relevance of a comprehensive and standardized validation of xCT calibration protocols.
Abstract
BACKGROUND AND PURPOSE: Accurate estimation of relative stopping power (RSP) is essential for proton therapy (PT) treatment planning. Emerging X-ray computed tomography (xCT) technologies are enabling more advanced RSP calibration strategies, but may also introduce greater variability across centers. This study aimed to assess inter-center variability in xCT calibration and RSP estimation for proton treatment planning, using a proton CT (pCT) prototype as reference. MATERIAL AND METHODS: An ex vivo phantom was transported to five European PT centers and one research center, and scanned with both single-energy and dual-energy xCT systems. Acquisitions were performed in three phantom configurations, simulating different anatomical sizes. The reconstructed xCT images were calibrated to RSP according to each center's clinical protocol and compared against the pCT-derived RSP maps. RESULTS: Voxel-by-voxel comparison between xCT-based and pCT-based RSP maps showed inter-center RSP variability of about 1% in soft tissues and up to 5% in bone, with larger differences observed for larger phantom sizes. The largest discrepancies (up to 5%) relative to pCT were observed in the same configurations. Centers implementing beam-hardening corrections generally showed closer agreement with pCT, whereas the effect of size-specific HLUTs was tissue- and anatomy-dependent. The impact of these RSP variations on water-equivalent thickness in representative clinical scenarios resulted in a 2σ inter-center variability of 1.09% and 1.39% for brain and prostate cases, respectively. CONCLUSIONS: This work presents the first inter-center comparison of xCT-based RSP estimation using an ex vivo phantom and pCT measurements as reference. The observed variability highlights the relevance of a comprehensive and standardized validation of xCT calibration protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of beam hardening on proton stopping-power ratio estimation using photon-counting, dual-energy, and single-energy computed tomography.Physics and imaging in radiation oncology · 2026
- Patient-Specific Stopping Power Calibration of Treatment Planning X-Ray CT Using Integrated Depth-Dose Profiles Acquired by Proton Transmission Imaging - A Feasibility Study.Physics in medicine and biology · 2026
- Evaluation of a fast kV switching dual energy CT in deriving relative stopping power using tissue equivalent phantoms of various sizesMedical Physics · 2026
- Proton treatment-plan dose verification with a DPM-based Monte Carlo engine.Frontiers in oncology · 2026
- An experimental imaging dose estimation method for the phase-II proton CT scanner.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Commissioning a single photon counting CT simulator for radiotherapy planning: Methods, challenges, and initial results.Journal of applied clinical medical physics · 2026