Impact of beam hardening on proton stopping-power ratio estimation using photon-counting, dual-energy, and single-energy computed tomography.
Lasse Bassermann, Jens Morgenthaler Edmund, Weronika Elżbieta Olech, Stine Elleberg Petersen, Maria Fuglsang Jensen, Ludvig Paul Muren, Vicki Trier Taasti
Danish Centre for Particle Therapy, Aarhus University Hospital, Aarhus, Denmark. · Department of Oncology, Herlev and Gentofte University Hospital, Herlev, Denmark.
Key points
From the paper's conclusion- SPR estimated based on PCCT-VMI with a HLUT showed reduced impact of beam hardening effects related to size and position.
Abstract
BACKGROUND AND PURPOSE: Accurate stopping-power ratio (SPR) estimation is essential for treatment planning of proton therapy. However, SPR accuracy is influenced by object size and position due to beam hardening. This study compared proton SPR estimation using photon-counting computed tomography (PCCT), dual-energy CT (DECT), and single-energy CT (SECT) across varying phantom sizes and insert positions. MATERIALS AND METHODS: Two phantoms with tissue-equivalent inserts were scanned using PCCT, DECT, and SECT scanners. Based on the PCCT and DECT scans, virtual monoenergetic images (VMIs) were created. Two SPR estimation methods were used: a DECT-based method and a SECT-based Hounsfield look-up table (HLUT). SPRs were calibrated on the Gammex Advanced Electron Density phantom and evaluated on a cylindrical phantom with four diameters. To assess size-dependency of the estimated SPR, tissue-equivalent bone inserts were placed centrally in the evaluation phantom, for each of the four diameters. To assess position-dependent SPR uncertainty, the inserts were placed at different distances from the centre of the evaluation phantom with fixed diameters. RESULTS: For smaller phantom diameters, SPR estimation accuracy was comparable across PCCT, DECT, and SECT. For larger phantom diameters (30 and 40 cm), SPRs estimated based on SECT and DECT showed position-dependency. PCCT demonstrated the most stable SPR estimations across positions, with deviations between 0.9% and 2.1%. SPR estimations were more stable using a HLUT than the DECT-based SPR method. CONCLUSIONS: SPR estimated based on PCCT-VMI with a HLUT showed reduced impact of beam hardening effects related to size and position.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-center comparison of relative stopping power estimation of an ex vivo phantom using clinical X-ray CT scanners and a proton CT prototype.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Impact of dual-layer spectral CT-derived stopping power ratio on single-energy CT-based carbon-ion treatment planning for thoracic cancer.PloS one · 2026
- An experimental imaging dose estimation method for the phase-II proton CT scanner.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Evaluation of a fast kV switching dual energy CT in deriving relative stopping power using tissue equivalent phantoms of various sizesMedical Physics · 2026
- Commissioning a single photon counting CT simulator for radiotherapy planning: Methods, challenges, and initial results.Journal of applied clinical medical physics · 2026
- Photon-counting CT versus Energy-integrating Detector CT Performance for Various BMI and Tumor Sizes in Lung CancerRadiology · 2026