Radiography (London, England : 1995)

Reducing radiographic retake rates with ultra-low dose pre-shots using deep learning-based noise reduction: A prospective clinical study.

K Mori, R Sekiguchi, M Suzaki, K Matsukura, Y Kidokoro, M Toukai, R Tsujimoto, T Negishi

Department of Radiological Technology, Saiseikai Kawaguchi General Hospital, 5-11-5 Nishikawaguchi, Kawaguchi, Saitama, 332-8558, Japan. Electronic address: k-mori@saiseikai.gr.jp. · Department of Radiological Technology, Saiseikai Kawaguchi General Hospital, 5-11-5 Nishikawaguchi, Kawaguchi, Saitama, 332-8558, Japan.

Observational · n=542

Key points

From the paper's conclusion
  • AiCE-enabled ultra-low-dose pre-shot imaging dramatically reduced retake rates while requiring minimal additional radiation dose.

Abstract

INTRODUCTION: The pre-shot technique is an effective approach to lower retake radiography rates; however, it is limited by increased patient radiation exposure. This study aimed to evaluate the impact of an ultra-low-dose pre-shot enabled by deep learning-based noise reduction (Advanced Intelligent Clear-IQ Engine: AiCE) on reducing retake rates in wrist radiography. METHODS: A CXDI-810C (Canon Medical Systems) was used as the X-ray detector. Phantom images simulating lateral wrist pre-shots with and without AiCE were acquired under fixed imaging conditions using an additional 1.0 mm copper filter. The tube current-time product was reduced from 2.8 mAs to 0.5 mAs. Fifteen radiological technologists determined the optimal imaging condition using a five-point confidence scale, with scores below three considered insufficient for retake decision-making. A prospective clinical evaluation was then conducted, comparing retake rates between a non-pre-shot group (n = 542) and a pre-shot group (n = 325). The entrance surface air kerma (Ka,e) was estimated from phantom measurements. RESULTS: Without AiCE, no imaging condition met the threshold required for retake decision-making. Conversely, with AiCE, a tube current-time product of 1.6 mAs achieved acceptable image quality (3.29 ± 0.47), with all evaluators assigning scores ≥3; therefore, it was selected as the optimal condition. In the clinical evaluation, the retake rate markedly decreased from 14.38% to 0.61%. The pre-shot Ka,e was 0.11 ± 0.03 μGy, corresponding to only 0.15% of the standard lateral wrist radiography dose. CONCLUSION: AiCE-enabled ultra-low-dose pre-shot imaging dramatically reduced retake rates while requiring minimal additional radiation dose. IMPLICATIONS FOR PRACTICE: The ultra-low-dose pre-shot using AiCE represents a highly effective strategy to minimize retake imaging with negligible dose penalty, offering substantial potential for dose optimization and improved workflow efficiency in clinical radiography.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers