Deep learning reconstruction for low‐dose thoracic CT of calcification inserts: A phantom study of CNR, noise magnitude, and noise texture
Ya‐Yun Hsiao, Cheng‐Chang Lu, Hui‐Chuan Kao, Tou‐Rong Chen
Key points
From the paper's conclusion- Purpose To compare filtered back projection (FBP), 50% Adaptive Statistical Iterative Reconstruction‐Veo (ASiR), and medium‐strength deep learning image reconstruction (DLIR) for hydroxyapatite calcification inserts under different dose, tube‐voltage, and phantom‐size conditions using contrast‐to‐noise ratio (CNR), background noise magnitude, and noise power spectrum (NPS) analysis.
Abstract
Purpose To compare filtered back projection (FBP), 50% Adaptive Statistical Iterative Reconstruction‐Veo (ASiR), and medium‐strength deep learning image reconstruction (DLIR) for hydroxyapatite calcification inserts under different dose, tube‐voltage, and phantom‐size conditions using contrast‐to‐noise ratio (CNR), background noise magnitude, and noise power spectrum (NPS) analysis. Methods A QRM‐Thorax containing a QRM‐CCI D100 cardiac calcification insert was scanned on a GE Revolution Apex CT system using institution‐defined routine and low‐dose protocols at 80, 100, 120, and 140 kVp, with and without an Extension Ring L. Images were reconstructed at 2.5‐mm thickness using FBP, ASiR, and DLIR. CNR was evaluated for 3‐ and 5‐mm inserts at 200, 400, and 800 mg HA/cm 3 . Background noise was assessed from a calcification‐excluded mask. For NPS analysis, 10 calcification‐free images were analyzed for each acquisition condition. Total two‐dimensional NPS power, power‐weighted normalized radial NPS, and power‐weighted mean spatial frequency were calculated. Results DLIR reduced background noise magnitude and increased CNR relative to FBP across the evaluated conditions, whereas differences relative to ASiR varied with insert size, density, tube voltage, and phantom size. Across all 16 protocol, phantom‐size, and tube‐voltage combinations, ASiR and DLIR reduced total 2D NPS power by 58.1% and 75.8%, respectively, relative to FBP; DLIR reduced total power by 41.9% relative to ASiR. ASiR produced a more consistent shift toward lower power‐weighted mean spatial frequencies, averaging 12.4% below FBP, whereas the shift with DLIR was smaller and more condition dependent, averaging 3.2% below FBP. Conclusions In this controlled phantom study, DLIR reduced total noise power and improved CNR relative to FBP, but its incremental difference from ASiR was task and protocol dependent. CNR and noise magnitude should be interpreted together with NPS because nonlinear reconstruction methods can alter both noise magnitude and spatial‐frequency distribution.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic PhantomarXiv · 2026
- Deep Learning-Based Reconstruction in Ultra-Low-Dose CT of the Ankle and Foot: A Comparative Study with Hybrid Iterative Reconstruction.Tomography (Ann Arbor, Mich.) · 2026
- Quantitative comparison of 68-keV monoenergetic, routine-dose, and low-dose 120-kVp CT for calcium-mimicking inserts in thoracic phantoms: effects of phantom size and reconstruction method.European radiology experimental · 2026
- Feasibility of deep learning-based image reconstruction using TrueFidelity for coronary calcium scoring.PloS one · 2026
- Deep-learning image reconstruction algorithms for CT: A task-based image quality assessment of four CT systems using a phantomDiagnostic and Interventional Imaging · 2026
- Quantitative and task‐based performance of spectral deep learning reconstruction in low‐dose dual‐layer coronary CT angiography: A phantom studyMedical Physics · 2026