Feasibility of AI-Denoised Ultra-Low-Dose CT for Detection of Acute Pelvic and Hip Fractures: A Pilot Multi-Reader Study.
Daniel Nguyen, Tina Shiang, Connie Ge, George Watts, Christopher Sereni, David Radcliffe, Hemang Kotecha, Gabriela Santos Nunez,
Department of Medicine, University of Massachusetts Chan Medical School, Worcester, 01655 MA, USA. · Department of Radiology, University of Massachusetts Chan Medical School, Worcester, 01655 MA, USA.
Key points
From the paper's conclusion- Conclusions: DLD-enhanced ULD-CT showed diagnostic accuracy and sufficiency similar to full-dose CT for pelvic and hip fracture detection, at a fraction of the radiation dose, though full-dose CT was rated higher in perceived image quality and diagnostic confidence and outperformed denoised low-dose CT numerically in hard cases.
Abstract
Objective: To evaluate the feasibility of simulated ultra-low-dose CT (ULD-CT) enhanced with deep learning denoising (DLD) for detecting pelvic and hip fractures. Method: This Institutional Review Board-approved retrospective study was performed at a multisite, single academic institution between October 2020 and May 2021, including 30 full-dose CT (CTfull) exams of acute pelvic and hip trauma (10 normal, 10 "easy," 10 "hard" cases by fracture conspicuity). Simulated low-dose CT images at 10% (CT10sim) and 5% (CT5sim) radiation dose were generated using a noise model-based algorithm and reconstructed from full-dose CT datasets. A commercial deep learning denoising algorithm enhanced these images, resulting in denoised low-dose CTs (CT10dld and CT5dld) for evaluation. The 90 resulting CT cases were reviewed independently for diagnostic accuracy, confidence, sufficiency, and imaging quality. Result: Diagnostic accuracy was similar across CT modalities overall (CTfull 80.8% [72.9, 86.9]; CT10dld 78.3% [70.1, 84.8]; CT5dld 77.5% [69.2, 84.1]) and within the easy and normal strata; in hard cases, accuracy was numerically lower for CT10dld (45.0%) and CT5dld (42.5%) than for CTfull (57.5%). Diagnostic sufficiency was similarly high across modalities (CTfull 91.7%, CT10dld 87.5%, CT5dld 88.3%). CT full was rated highest in perceived image quality across all strata (mean 4.21 ± 0.83 overall vs. 3.47 ± 0.97 for CT10dld and 3.23 ± 0.98 for CT5dld) and in diagnostic confidence (mean 4.61 ± 0.63 vs. 4.45 ± 0.71 and 4.41 ± 0.72, respectively), though neither translated into a corresponding drop in diagnostic accuracy overall. Inter-reader agreement was highest between MSK radiologists, reaching substantial agreement on CT10dld (κ = 0.760) and CT5dld (κ = 0.672). Conclusions: DLD-enhanced ULD-CT showed diagnostic accuracy and sufficiency similar to full-dose CT for pelvic and hip fracture detection, at a fraction of the radiation dose, though full-dose CT was rated higher in perceived image quality and diagnostic confidence and outperformed denoised low-dose CT numerically in hard cases. Full-dose CT remains the preferred modality for complex and high-energy trauma, while denoised low-dose CT modalities represent a promising option for straightforward presentations where radiation minimization is a priority.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Deep Learning-Based Reconstruction in Ultra-Low-Dose CT of the Ankle and Foot: A Comparative Study with Hybrid Iterative Reconstruction.Tomography (Ann Arbor, Mich.) · 2026
- Automatic rib fracture detection on postmortem CT data using deep learningInternational Journal of Legal Medicine · 2025
- Deep learning reconstruction for low‐dose thoracic CT of calcification inserts: A phantom study of CNR, noise magnitude, and noise textureJournal of Applied Clinical Medical Physics · 2026
- Deep-learning image reconstruction algorithms for CT: A task-based image quality assessment of four CT systems using a phantomDiagnostic and Interventional Imaging · 2026
- Performance of artificial intelligence for fracture detection: an umbrella review and second-order-meta-analysisnpj Digital Medicine · 2026
- Independent bone-level diagnostic accuracy study of an AI tool for detecting appendicular skeletal fractures on radiographsEuropean Radiology · 2026