Evaluation of a simplified immobilization approach in thoracic SBRT: A comparative analysis of two patient positioning systems.
C Valenta, B Colson, M Mollet, J Billard, L Ettalhaoui, F Felici, C Invernizzi, S Guendouzen,
Department of Radiation Oncology, Institut Godinot, Reims, France. · Department of Radiation Oncology, Institut Godinot, Reims, France; Université de Reims Champagne-Ardenne, CRESTIC, Reims, France.
Key points
From the paper's conclusion- Both immobilization systems provided adequate positioning reproducibility for thoracic SBRT. Lower z-axis displacement was observed with the thermoformed chin-chest mask, although respiratory insufficiency may have contributed to this difference.
Abstract
INTRODUCTION: The choice of immobilization system in thoracic stereotactic body radiation therapy (SBRT) influences patient comfort, workflow efficiency, and position reproducibility. This study compared the reproducibility of two immobilization strategies, using a vacuum mattress alone versus a reference immobilization setup combining a vacuum mattress with a thermoformed chin-chest mask for thoracic SBRT. METHODS: This retrospective study included 16 patients treated with thoracic SBRT: 11 positioned with a vacuum mattress combined with a thermoformed chin-chest mask (control group) and 5 with a vacuum mattress alone (investigational group). All patients underwent 4DCT simulation under free-breathing conditions to assess tumor motion. Translational displacements (left-right [x], superior-inferior [y], anterior-posterior [z]) and rotational shifts (pitch, yaw, roll) were recorded over 71 treatment fractions. Statistical comparisons between the two immobilization groups were performed for patient characteristics and outcome measures using Fisher's exact test for categorical variables and Wilcoxon rank-sum tests. RESULTS: Among the 71 fractions, 70.4% were delivered using the mask-and-mattress system and 29.6% using the mattress-only setup. Patient characteristics, including age, tumor location, and body mass index, were retrospectively collected from medical records and were comparable between groups. Mean translational (0.4 cm) and rotational (1.2°) displacements were similar between groups. Displacement along the z-axis was significantly greater in the investigational group (p = 0.008), particularly for lower-lobe tumors (p = 0.005). CONCLUSION: Both immobilization systems provided adequate positioning reproducibility for thoracic SBRT. Lower z-axis displacement was observed with the thermoformed chin-chest mask, although respiratory insufficiency may have contributed to this difference. IMPLICATIONS FOR PRACTICE: A vacuum mattress alone may represent a simpler alternative for thoracic SBRT in appropriately selected patients, while additional caution may be warranted in patients with respiratory insufficiency.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Beyond CBCT: real-time motion management in spine and non-spine bone SBRT using stereoscopic X-rays and surface guidance.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Maskless head and neck radiotherapy using surface-guided radiation therapy: initial clinical feasibility of real-time setup, intrafraction monitoring, and automatic beam-hold.Journal of applied clinical medical physics · 2026
- The Prospective SPOTLESS Trial: Setup Accuracy of Tattoo-Less Surface-Guided Breast Radiotherapy Including Regional Nodal Irradiation.Journal of clinical medicine · 2026
- Geometric Motion of Multiple Liver Targets in Stereotactic Body Radiotherapy: Analysis of Translational and Rotational Motion with Different Motion-Management and Isocenter techniques.Biomedical physics & engineering express · 2026
- Maintaining setup reproducibility in patients unable to achieve bilateral arm abduction during breast radiotherapy: evaluation of a novel arm support deviceJournal of Radiotherapy in Practice · 2026
- Surface-guided versus conventional setup in rectal cancer radiotherapy: A workflow-level comparison of CBCT-Derived setup corrections.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026