An EPID-based virtual audit framework for multi-institutional radiotherapy audits using direct fluence comparison.

Maiko Kishigami, Takanori Adachi, Yoshitomo Ishihara, Hiroyuki Ito, Tomohiro Ono, Yukako Kishigami, Masahide Saito, Masato Tsuneda,

Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, 53 Kawahara-cho, Shogoin, Sakyo-ku, Kyoto 606-8507, Japan. · Department of Radiation Oncology and Image-Applied Therapy, Graduate School of Medicine, Kyoto University, 54 Kawahara-cho, Shogoin, Sakyo-ku, Kyoto 606-8507, Japan.

Key points

From the paper's conclusion
  • Leave-one-machine-out validation supported the feasibility of vendor-level parameter standardization, potentially reducing machine-specific optimization. A preliminary tiered framework combining 95% at 3%/1 mm and 90% at 3%/2 mm may identify cases requiring review in multi-institutional radiotherapy audits.

Abstract

BACKGROUND AND PURPOSE: The complexity of volumetric modulated arc therapy highlights the need for robust quality assurance in multi-institutional clinical trials. Although electronic portal imaging device-based virtual audits are scalable, standardized workflows and evaluation criteria remain lacking. This study developed and evaluated a virtual audit framework comparing radiotherapy plan-derived and electronic portal imaging device-derived fluence maps. MATERIAL AND METHODS: Forty-one volumetric modulated arc therapy datasets from 15 institutions were analyzed. Plan-derived fluence maps were generated by convolving opening density matrices with a double-Gaussian electronic portal imaging device response kernel, while integrated portal images served as detector-derived fluence maps. Vendor-specific common parameters were evaluated using leave-one-machine-out validation, excluding each held-out machine's parameters before recalculating vendor-level parameters. Multileaf collimator positional, collimator-angle, and monitor-unit errors were evaluated using receiver operating characteristic analysis. RESULTS: Mean gamma passing-rate differences ranged from -0.015 to 0.018 percentage points for Varian systems and from 0.387 to 1.018 percentage points for Elekta systems. The 3%/1 mm criterion yielded areas under the receiver operating characteristic curves of 0.959, 0.938, and 0.898 for multileaf collimator positional, collimator-angle, and monitor-unit errors, respectively, and produced the highest values for multileaf collimator positional and collimator-angle errors. A 95% gamma passing rate threshold at 3%/1 mm provided high sensitivity, whereas 90% at 3%/2 mm provided high specificity. CONCLUSIONS: Leave-one-machine-out validation supported the feasibility of vendor-level parameter standardization, potentially reducing machine-specific optimization. A preliminary tiered framework combining 95% at 3%/1 mm and 90% at 3%/2 mm may identify cases requiring review in multi-institutional radiotherapy audits.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers