Prediction of patient-specific quality assurance for volumetric modulated arc therapy using machine log files.
Yuhan Fan, Jiawen Shang, Peng Huang, Zhihui Hu, Ke Zhang, Xin Xie, Zhiqiang Liu, Hui Yan,
Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China. · State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Disease, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Key points
From the paper's conclusion- The proposed model that combines beam intensity map and MLC leaf positional error map exhibited a better prediction performance for GPRs of VMAT PSQA. It would be advantageous to include more motion precision information of treatment machine into DL models for high-performance prediction of VMAT PSQA.
Abstract
BACKGROUND: Deep learning (DL) models have been used for predicting gamma passing rates (GPRs) of patient specific quality assurance (PSQA) in volumetric modulated arc therapy (VMAT). As the model input mostly from planning parameters and dose distribution, the effect of motion precision of the treatment machine has frequently been ignored. In this study,the most recent motion precision of multi-leaf collimator (MLC) leaves was estimated by their records in machine log file and used jointly with the beam intensity map to predict GPRs of VMAT PSQA. METHODS: The input of the DL model consists of two parts: the MLC positional error map which was estimated from machine log file and the beam intensity map which was obtained from the treatment plan. A UNet++ model was trained to predict the GPR of VMAT PSQA. The effect of feature maps and time intervals in estimating MLC leaf positional error maps on the prediction accuracy was evaluated. In addition, the proposed model was compared with two competing DL models (ResNet and SE-Net) and a classic regression model. The prediction error was quantified by mean absolute errors (MAEs) under two clinical accepted criteria (3%-3 mm and 3%-2 mm). Pearson's correlation coefficient (r) between measured and predicted GPR was also assessed. RESULTS: The model with the inputs of beam intensity and MLC leaf positional error maps achieved the best performance among all combinations with various input feature maps. The MAEs predicted by the proposed model with MLC leaf positional error map obtained from 5 days' records were 0.90 for 3%-3 mm criterion and 1.74 for 3%-2 mm criterion, whereas those predicted by the proposed model with MLC leaf positional error map obtained from 1 day's record were 1.01 for 3%-3 mm criterion and 1.85 for 3%-2 mm criterion. The proposed model achieved the smaller MAEs (1.01 for 3%-3 mm criterion and 1.85 for 3%-2 mm criterion) than the other two competing models: ResNet (2.85 for 3%-3 mm criterion and 2.83 for 3%-2 mm criterion) and SE-Net (3.53 for 3%-3 mm criterion and 4.55 for 3%-2 mm criterion). CONCLUSIONS: The proposed model that combines beam intensity map and MLC leaf positional error map exhibited a better prediction performance for GPRs of VMAT PSQA. It would be advantageous to include more motion precision information of treatment machine into DL models for high-performance prediction of VMAT PSQA.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Predicting VMAT patient-specific quality assurance outcome using a machine learning workflow.Physical and engineering sciences in medicine · 2026
- Machine learning-based prediction of gamma passing rates in VMAT prostate external radiation therapy plans.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Daily Multi-Leaf Collimator positional error prediction using machine learning: A Dynalog-based approach for patient specific quality assurance.Journal of applied clinical medical physics · 2026
- General versus site specific quantile random forest models for virtual patient-specific quality assurance of VMAT plans.Journal of applied clinical medical physics · 2026
- Machine learning prediction of electronic-portal-imaging-device-based in vivo dosimetry results using plan complexity metrics, machine output and gamma passing rate: a feasibility study.Physical and engineering sciences in medicine · 2026
- Deep learning for gamma map prediction in breast IMRT verificationMedical Physics · 2026