Effect of right ventricular outflow tract material properties on simulated transcatheter pulmonary placement.
Jalaj Maheshwari, Wensi Wu, Christopher N Zelonis, Steve A Maas, Kyle Sunderland, Yuval Barak-Corren, Stephen Ching, Patricia Sabin,
Department of Anesthesiology and Critical Care Medicine, Children's Hospital of Philadelphia, Philadelphia, 19104, PA, USA. · Cardiovascular Institute, Children's Hospital of Philadelphia, Philadelphia, 19104, PA, USA; Department of Mechanical Engineering and Applied Mechanics, University of Pennsylvania, Philadelphia, 19104, PA, USA; Division of Pediatric Cardiology, Children's Hospital of Philadelphia, Philadelphia, 19104, PA, USA.
Key points
From the paper's conclusion- Finite element (FE) simulations emulating transcatheter pulmonary valve (TPV) system deployment in patient-specific right ventricular outflow tracts (RVOT) assume material properties for the RVOT and adjacent tissues.
Abstract
Finite element (FE) simulations emulating transcatheter pulmonary valve (TPV) system deployment in patient-specific right ventricular outflow tracts (RVOT) assume material properties for the RVOT and adjacent tissues. Sensitivity of the deployment to variation in RVOT material properties is unknown. Moreover, the effect of a transannular patch stiffness and location on simulated TPV deployment has not been explored. A sensitivity analysis on the material properties of a patient-specific RVOT during TPV deployment, modeled as an uncoupled HGO material, was conducted using FEBioUncertainSCI. Further, the effects of a transannular patch during TPV deployment were analyzed by considering two patch locations and four patch stiffnesses. Visualization of results and quantification were performed using custom metrics implemented in SlicerHeart and FEBio. Sensitivity analysis revealed that the shear modulus of the ground matrix (c), fiber modulus (k1), fiber dispersion (κ), and fiber mean orientation angle (γ) had the greatest effect on 95th %ile stress, whereas only c had the greatest effect on 95th %ile Lagrangian strain. First-order sensitivity indices contributed the greatest to the total-order sensitivity indices. Simulations using a transannular patch revealed that peak stress and strain were dependent on patch location and the underlying vessel geometry, with increasing patch stiffness modulating the magnitude of local stress concentrations. As stiffness of the patch increased, greater stress was observed at the interface connecting the patch to the RVOT, and stress in the patch itself increased while strain decreased. The total enclosed volume by the TPV device remained unchanged across all simulated patch cases. This study highlights that while uncertainties in tissue material properties and patch locations may influence functional outcomes, FE simulations provide a reliable framework for evaluating these outcomes in TPVR.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In Silico Insights into Device-Artery Interactions in Endovascular Pulmonary Flow Restrictors for Neonates with Congenital Heart Defects.Cardiovascular engineering and technology · 2026
- Impact of Leaflet Thickness on Polymeric Valve Scaffold Function and Remodeling: Assessment in a 1- and 3-Month Ovine Stented Pulmonary Valve Replacement Model.Journal of biomedical materials research. Part A · 2026
- Experimental Bench Validation of a Computational Fluid-Structure Interaction Model of Transcatheter Aortic Valve DynamicsarXiv · 2026
- On analytical evaluation of radial and bending stiffness of transcatheter aortic valve stent lattice structureFrontiers in Materials · 2026
- Learning from Biodegradable Coronary Stents: Future Directions for TPVR Biodegradable Stents.International journal of molecular sciences · 2026
- Patient-specific Modeling of the Tricuspid Valve from 3D Transesophageal EchocardiographyarXiv · 2026