In Silico Insights into Device-Artery Interactions in Endovascular Pulmonary Flow Restrictors for Neonates with Congenital Heart Defects.
Sara Bonfanti, Maria Alexandra Cetatoiu, Paola Di Vincenzo, Emanuele Gasparotti, Benigno Marco Fanni, Shannen B Kizilski, Dominic P Recco, Pedro Del Nido,
Laboratory of Biological Structure Mechanics (LaBS), Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Piazza Leonardo da Vinci 32, 20133, Milano, Italy. · BioCardioLab, Bioengineering Unit-Heart Hospital, Fondazione Toscana Gabriele Monasterio, Massa, Italy.
Key points
From the paper's conclusion- This study demonstrates the feasibility of a patient-specific computational framework for simulating the deployment of endovascular pulmonary flow restrictors using a validated MVP™ digital twin.
Abstract
INTRODUCTION: Transcatheter pulmonary flow restriction using a modified Medtronic Vascular Plug (MVP™) has recently emerged as a promising alternative to surgical pulmonary arterial banding in neonates with complex congenital heart defects. However, clinical outcomes remain variable, and the underlying biomechanical mechanisms are poorly understood. This study aims to exploit in silico modeling to improve the understanding of device-artery interactions following the deployment of modified MVP™ in neonatal pulmonary arteries. To this end, we (i) developed and validated a high-fidelity digital twin of an MVP™ device, and (ii) used clinical imaging data from a retrospective cohort of seven neonates treated at Boston Children's Hospital to reproduce in silico the in vivo device-artery interaction in both the right and left pulmonary arteries. METHODS: The digital twin was reconstructed from high-resolution optical measurements and mechanically characterized under both local and global compression. Idealized artery geometries were generated based on pre-procedural angiographic measurements. For each patient, wall thickness was varied within a size-adjusted range derived from the literature, while Young's modulus was calibrated to reproduce the post-implantation arterial diameters, thereby quantifying the vessel's stiffness in response to device expansion. RESULTS: Case-specific simulations revealed marked variability in pulmonary artery effective stiffness across patients and between arterial branches. While identical post-implantation diameters could be reproduced through different thickness-stiffness combinations, these generated substantially different local contact pressure distributions. CONCLUSIONS: This study demonstrates the feasibility of a patient-specific computational framework for simulating the deployment of endovascular pulmonary flow restrictors using a validated MVP™ digital twin.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Outcomes of Modified Microvascular Plugs as a Pulmonary Flow Restrictors in Congenital Heart Disease.Pediatric cardiology · 2026
- In Vitro Hemodynamic Assessment of a Novel Pulmonary Visceral Pleura (PVP) Transcatheter Aortic Valve Replacement Prototype Based on Optical and Acoustic Measurements.Bioengineering (Basel, Switzerland) · 2026
- Preconditioned Endothelialized Grafts Reduce Stenosis and Improve Vascular Remodeling in a Rat Pulmonary Artery Model.Acta biomaterialia · 2026
- Feasibility study of fluid-structure-growth modelling in the context of endogenous tissue restorationbioRxiv · 2026
- A biomimetic, bioengineered polymeric pulmonary valve promotes native tissue remodeling and thrombosis-free function in a chronic ovine modelEuropean Heart Journal · 2026
- Learning from Biodegradable Coronary Stents: Future Directions for TPVR Biodegradable Stents.International journal of molecular sciences · 2026