Frontiers in Materials

On analytical evaluation of radial and bending stiffness of transcatheter aortic valve stent lattice structure

Alex G. Kuchumov, Gleb Selivanov

Key points

From the paper's conclusion
  • A useful application of this theoretical approach is additive manufacturing of these stents by adjusting stent design to meet particular clinical and anatomical needs to find a balance between radial stiffness, bending flexibility, and structural performance.

Abstract

Background and Objective Mechanical performance of transcatheter aortic valve stents (radial and bending stiffness) primarily depends on their geometry and cell architecture. Finite element simulations and experimental testing of these cellular lattice structures are computationally expensive and unsuitable for quick design optimization, especially for patient-specific devices, which can be additively manufactured. In order to predict the radial and bending stiffness of TAVI stents directly from the parametric lattice shape, an analytical framework must be developed. Methods The TAVI frame was represented parametrically. A periodic diamond cellular lattice made up of circular struts was used to depict the stent. The elastic strain energy related to strut bending and axial deformation was calculated using Castigliano’s second theorem and Euler-Bernoulli beam theory. In order to produce an equal bending stiffness for global bending, the strain energy that resulted from projecting the macroscopic curvature onto the local deformation of each lattice cell/strut was homogenized. The force-displacement response for compression between parallel rigid plates was obtained from the derivative of the total strain energy with respect to the prescribed displacement, and the imposed plate displacement was connected to the deformation of the diamond lattice. TAVI lattice specimens were mechanically tested to compare the analytical predictions. Results The TAVI lattice geometry and its global mechanical response were found to be explicitly related by the analytical formulation. Strut diameter, lattice topology, circumferential cell number, axial cell number, and local strut length all significantly influenced the anticipated bending stiffness; the bending-dominated contribution scaled with the fourth power of strut diameter. The effects of lattice bending and axial strut deformation to the force-displacement response were also included in the plate-compression equation. The main patterns seen in the experimental observations were replicated by the analytical framework. Conclusion The proposed analytical approach can be adopted for geometry optimization and patient-specific TAVI frame design. A useful application of this theoretical approach is additive manufacturing of these stents by adjusting stent design to meet particular clinical and anatomical needs to find a balance between radial stiffness, bending flexibility, and structural performance.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers