Selecting a spectrometric detector for drinking water monitoring in the case of nuclear and radiological emergencies.
Jonathan Sundström, Charlotta Nilsson, Mats Isaksson, Christopher Rääf
University of Gothenburg, Department of Medical Radiation Sciences, Sweden. Electronic address: jonathan.sundstrom@gu.se. · Lund University, Department of Physics, Sweden.
Key points
From the paper's conclusion- The results show that both detection efficiency and energy resolution must be considered when selecting detector systems for radionuclide-specific drinking-water monitoring following an RN emergency.
Abstract
Rapid and reliable radionuclide-specific monitoring of drinking water may be required following a radiological or nuclear (RN) emergency. In this study, ten gamma-detector configurations were evaluated for continuous drinking-water monitoring using Monte Carlo simulations combined with measured detector backgrounds. Two source terms were investigated, representing fallout from a nuclear weapon detonation and a nuclear power plant accident, each at 48 h and 7 d after release. Two measurement geometries were simulated: a 4 L Marinelli beaker and a shielded detector positioned directly against a water pipe. A conservative scenario was assumed in which the complete source-term radionuclide composition was transferred to the drinking water. Minimum Detectable Activity (MDA) and Minimum Quantifiable Activity (MQA) were calculated for 131I and 137Cs, including contributions from other radionuclides within the respective regions of interest. Larger detectors generally provided lower MDA and MQA values, with the lowest values mainly obtained for HPGe. Despite substantially lower volume-normalised full-energy peak efficiency for the Pipe geometry, its concentration-based MDA and MQA values were relatively similar to those of the Marinelli geometry because of the much larger sampled water volume. Energy resolution strongly affected radionuclide-specific quantification. For 131I, 7 d after a nuclear weapon detonation in a Marinelli geometry, non-target radionuclides contributed approximately 79% of the NaI(Tl) ROI counts, compared with approximately 29% for HPGe. The results show that both detection efficiency and energy resolution must be considered when selecting detector systems for radionuclide-specific drinking-water monitoring following an RN emergency.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of high-purity germanium detectors and gamma-ray water monitors in radioactive wastewater management for targeted radionuclide therapy.Annals of nuclear medicine · 2026
- Urban Dose Rate Estimation Using Distributed Static D3S Gamma-Ray Spectrometers.Sensors (Basel, Switzerland) · 2026
- Environmental radiological baseline characterization of a proposed medical cyclotron facility in Indonesia using dual-detector measurements and geostatistical analysis.Environmental geochemistry and health · 2026
- Application of a low-cost Geiger-Müller monitoring system for clinical radiation environments.Journal of applied clinical medical physics · 2026
- Optimizing radon quantification in arid groundwater: Counting-time strategies and QA/QC for sub-Bq·L-1 measurements.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Fast determination of americium and curium isotopes in urine samples and internal dose evaluation in emergency situationsJournal of Radioanalytical and Nuclear Chemistry · 2026