Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine

Optimizing radon quantification in arid groundwater: Counting-time strategies and QA/QC for sub-Bq·L-1 measurements.

Al Mamun, Abdullah Al-Mamun, Md Shafiquzzaman, Aljawad Mohammed Alolaywi, Maha Alruwaili, Amira Salman Alazmi

Department of Physics, College of Science, University of Hafr Al Batin, Al Jamiah, Hafr Al Batin, 39524, Saudi Arabia. Electronic address: almamun@uhb.edu.sa. · Department of Civil and Environmental Engineering, King Fahd University of Petroleum & Minerals, Dhahran, 31261, Saudi Arabia. Electronic address: mdabdullah.mamun@kfupm.edu.sa.

Guideline

Key points

From the paper's conclusion
  • We define "optimization" as selecting the minimal, decision-appropriate RAD7 Big-Bottle operating configuration that meets a target minimum detection limit/limit of quantification (MDL/LOQ) and precision should be achieved here by integrating (i) a tiered counting-time strategy (screening: 30-150 min; confirmatory: 900-1500 min; ultralow: ∼3000 min), (ii) temperature/salinity/loop-volume-aware air

Abstract

Low-level radon-in-water measurements in arid aquifers are constrained by counting statistics at short runtimes and by systematic air-water partition bias in closed-loop degassing systems. We define "optimization" as selecting the minimal, decision-appropriate RAD7 Big-Bottle operating configuration that meets a target minimum detection limit/limit of quantification (MDL/LOQ) and precision should be achieved here by integrating (i) a tiered counting-time strategy (screening: 30-150 min; confirmatory: 900-1500 min; ultralow: ∼3000 min), (ii) temperature/salinity/loop-volume-aware air-water conversion, and (iii) QA/QC gates (RH ≤ 7%, calibration acceptance, blanks/duplicates/drift criteria). Precision and Currie-based MDL/LOQ improve with integration time (t-1/2), with practical stability achieved at ∼900-1500 min for routine reporting. Raw air-phase readings showed a systematic enrichment (∼1.26×) relative to dissolved estimates; phase conversion removed this effect, yielding agreement diagnostics consistent with bias reduction. At long runtimes, uncertainty becomes bounded by fixed components (calibration and environmental inputs), indicating that further improvement requires tightening these terms rather than extending runtime alone. This workflow provides a traceable, decision-grade approach to sub-Bq·L-1 radon surveillance in arid groundwater monitoring programs.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts