Comparison of high-purity germanium detectors and gamma-ray water monitors in radioactive wastewater management for targeted radionuclide therapy.
Noriaki Miyaji, Yuma Ihara, Rintaro Sato, Shigenori Yabe, Katsuyuki Kikori, Naochika Akiya, Kaito Wachi, Akemi Minami,
Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, 10-6 Sakaemachi, Fukushima-Shi, Fukushima, 960-8516, Japan. noriaki.miyaji@gmail.com. · Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, 10-6 Sakaemachi, Fukushima-Shi, Fukushima, 960-8516, Japan.
Key points
From the paper's conclusion- Conversion factors reduced the gap between GWM and HPGe detection and provided estimates closer to HPGe-based values, facilitating more efficient discharge management in targeted radionuclide therapy facilities.
Abstract
OBJECTIVE: Gamma-ray water monitors (GWMs) are widely used for radioactive wastewater monitoring but can be inaccurate in mixed radionuclide conditions. This study compared radioactivity concentrations measured using a GWM and a high-purity germanium (HPGe) detector and evaluated the validity of derived conversion factors. METHODS: Wastewater containing 131I and 177Lu with 177mLu impurity was sampled from storage tanks in a targeted radionuclide therapy facility. We then measured radioactivity concentrations using GWM and HPGe detectors. GWM reproducibility was evaluated, and the effects of sampling location and stirring were assessed. Conversion factors were derived from paired measurements and applied to independent samples. RESULTS: Radioactivity concentrations did not differ significantly between sampling locations or before and after stirring. GWM measurements showed high reproducibility (CV = 1.7%) but consistently produced higher radioactivity concentrations than the HPGe detector. Summed concentration limit ratios were 30.9 and 3.0 for the GWM and the HPGe detectors, respectively, whereas a conversion factor-based estimate was 13.9. CONCLUSION: Conversion factors reduced the gap between GWM and HPGe detection and provided estimates closer to HPGe-based values, facilitating more efficient discharge management in targeted radionuclide therapy facilities.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Selecting a spectrometric detector for drinking water monitoring in the case of nuclear and radiological emergencies.Journal of environmental radioactivity · 2026
- Application of a low-cost Geiger-Müller monitoring system for clinical radiation environments.Journal of applied clinical medical physics · 2026
- Environmental radiological baseline characterization of a proposed medical cyclotron facility in Indonesia using dual-detector measurements and geostatistical analysis.Environmental geochemistry and health · 2026
- Evaluation of internal doses from liver, kidneys, and urinary bladder irradiation during radionuclide clearance: Extension of S-values dataset for reference adult male and female mesh-type phantoms.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Engineered Activated Carbons for Multi-Radionuclide Removal from Real Low-Level Radioactive Wastewater.Nanomaterials (Basel, Switzerland) · 2026