Pacing cessation just before generator replacement - A life-threatening consequence of pacemaker battery depletion: A case report.
Ryoto Mitsumizo, Yusuke Takahara, Hiroshi Ikuta, Ryota Komoto, Ryuichi Matsukawa, Keita Inanaga, Jun-Ichiro Nishi
Clinical Training Center, Aso Iizuka Hospital, Japan. · Department of Cardiology, Aso Iizuka Hospital, Japan.
Key points
From the paper's conclusion- We report the case of a 77-year-old man who underwent single-chamber pacemaker implantation 16 years prior.
Abstract
UNLABELLED: We report the case of a 77-year-old man who underwent single-chamber pacemaker implantation 16 years prior. The patient, who did not utilize remote monitoring, was lost to follow-up 14 years after the implantation, leading to profound battery depletion. This lack of surveillance resulted in a critical reduction of the pacing rate to 14 ppm, culminating in the complete cessation of pacing immediately before generator replacement. Despite the Abbott device's specified End-of-Life (EOL) safety mechanism, pacing function failed entirely-a rare but life-threatening complication. Uniquely, this case provides serial electrocardiographic documentation capturing the full progression of severe battery depletion. These findings underscore that continuous follow-up and remote monitoring, as emphasized in recent guidelines, are pivotal for preventing such critical outcomes. LEARNING OBJECTIVE: This case highlights the potential for profound pacemaker rate reduction and complete pacing cessation beyond manufacturer-defined End-of-Life settings in the setting of extreme battery depletion. It emphasizes the importance of regular device follow-up, timely generator replacement, and the use of remote monitoring to detect critical battery status changes and prevent life-threatening complications during pacemaker management.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A decade of leadless pacemaker implantation: technical performance, safety and device longevityEuropean Heart Journal · 2026
- Simple programming, significant impact: Pacemaker Sleep Function extends device longevityHeartRhythm Case Reports · 2026
- Leadless Pacing Dyssynchrony and Diastolic Mitral Regurgitation Rescued by Left Bundle Branch Area Pacing.JACC. Case reports · 2026
- Ten years follow-up data on a leadless pacing system – real life experience in comparison to data of the prospective trialEuropean Heart Journal · 2026
- Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation.JACC. Case reports · 2026
- Case Report: Electrode perforation with normal pacemaker parameters leading to hemorrhagic shock: a rare case report of the "penetration-exit-re-entry" phenomenon.Frontiers in medicine · 2026