Case Report: Electrode perforation with normal pacemaker parameters leading to hemorrhagic shock: a rare case report of the "penetration-exit-re-entry" phenomenon.
Zuoan Qin, Jian Yang, Li Luo, Shi Chen, Fei Yang
Department of Cardiology, Changde Hospital, Xiangya School of Medicine, Central South University (The First People's Hospital of Changde City), Changde, China. · Department of Cardiac Catheterization Laboratory, Changde Hospital, Xiangya School of Medicine, Central South University (The First People's Hospital of Changde City), Changde, China.
Key points
From the paper's conclusion- This case highlights that normal intraoperative pacing parameters cannot rule out occult extravascular lead perforation and venous laceration. Operators should maintain high vigilance for unexplained perioperative hemodynamic collapse. Combined imaging, diagnostic thoracentesis, and early thoracoscopic surgical intervention are the core rescue strategies for this fatal complication.
Abstract
BACKGROUND: Permanent pacemaker implantation via subclavian vein puncture may induce rare but life-threatening iatrogenic vascular injury. Vascular perforation with pacing lead penetration-exit-re-entry trajectory accompanied by normal intraoperative pacing parameters is reported extremely rarely, which greatly increases the risk of missed diagnosis. CASE PRESENTATION: A 70-year-old female patient with complete atrioventricular block received dual-chamber pacemaker implantation. All intraoperative lead electrical indicators were within normal range. Immediately after the operation, she rapidly developed massive hemothorax and hemorrhagic shock with severe hypoxemia. A thoracoscopic exploration verified a unique "penetration-exit-re-entry" lesion: the pacing electrode penetrated the left subclavian vein wall, exited the vascular lumen, and re-entered the innominate vein, resulting in persistent intrathoracic bleeding. After emergency thoracic drainage, surgical venous repair and blood transfusion resuscitation, the patient's vital signs gradually stabilized and fully recovered. A 1-month follow-up showed stable pacemaker function without adverse events. CONCLUSION: This case highlights that normal intraoperative pacing parameters cannot rule out occult extravascular lead perforation and venous laceration. Operators should maintain high vigilance for unexplained perioperative hemodynamic collapse. Combined imaging, diagnostic thoracentesis, and early thoracoscopic surgical intervention are the core rescue strategies for this fatal complication.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Superior vena cava perforation during atrial lead implantation in a patient with repaired tetralogy of Fallot: a case report.European heart journal. Case reports · 2026
- Contrasting ultrasound findings in pacemaker lead perforation with and without cardiac tamponade.BMJ case reports · 2026
- Delayed innominate vein perforation and massive hemothorax following left-sided PICC placement: A case report.Medicine · 2026
- Subacute Microperforation of Right Atrial Pacemaker Lead Presenting with Shock due to Cardiac Tamponade: A Case Report.Clinical practice and cases in emergency medicine · 2026
- A Perforation Mimic After Temporary Pacing Lead Migration: Surgical Confirmation of Marked Apical Wall Thinning.JACC. Case reports · 2026
- Intraprocedural tail-first pulmonary artery migration of a leadless pacemaker: a protective sleeve-assisted helix-side retrieval strategy-a case report.European heart journal. Case reports · 2026