Variability and Drivers of Battery Longevity in Contemporary Implantable Cardioverter-Defibrillator Systems Across Geographic Regions.

Niraj Varma, Alan Bulava, Eduardo Celentano, Amine Tazibet, Abhishek J Deshmukh, Tobias Timmel, Alessio Gargaro, Christian Ramharter,

Cleveland Clinic, Cleveland, Ohio, USA. Electronic address: varman@ccf.org. · University Hospital Olomouc and Faculty of Medicine, Palacky University, Olomouc, Czechia.

Observational

Key points

From the paper's conclusion
  • Battery longevity of contemporary ICD/CRT-D systems varies substantially across geographic regions despite identical device technology, and may be potentially extended by attention to pacing-related device use and follow-up methods.

Abstract

BACKGROUND: Battery longevity of implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) affects clinical outcomes and resource use, yet the contribution of regional practice patterns to battery consumption remains poorly characterized. OBJECTIVES: We compared projected battery longevity of contemporary ICD and CRT-D systems across Asia-Pacific, Europe, and North America and evaluated factors associated with regional differences. METHODS: We retrospectively analyzed 138,897 transvenous ICD/CRT-D devices (market release 2019) from 36 countries in the global BIOTRONIK Home Monitoring database. Longevity was estimated using a validated linear battery-drain projection model. Devices were grouped by type and geographic region. RESULTS: Median (Q1-Q3) longevity was highest for single-chamber ICD (14.8 [Q1-Q3: 14.0-15.5] years) and lowest for CRT-D (9.3 [Q1-Q3: 7.9-10.5] years). Longevity was longer in Asia-Pacific and Europe than in North America, with the longest estimates observed in Asia-Pacific subregional clusters and regional differences up to 2.0 years. High-voltage consumption showed limited regional variation. Larger differences were observed for low-voltage consumption and atrial pacing burden (8.7%-40.1%). Remote monitoring contributed <1% to battery drain, whereas telemetry-related energy consumption for in-office visits exhibited the greatest regional variability. Multivariable analyses identified pacing burden, follow-up intensity, and delivered shocks as independent correlates of projected battery longevity. Adjustment for these factors attenuated, but did not eliminate, regional differences. CONCLUSIONS: Battery longevity of contemporary ICD/CRT-D systems varies substantially across geographic regions despite identical device technology, and may be potentially extended by attention to pacing-related device use and follow-up methods.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers