Injury

Additional FiberTape-Cerclage® augmentation or "Bumper-plate" osteosynthesis for symphyseal arthrodesis - A biomechanical study.

Tobias Fritz, Jasmin Keßler, Marcel Orth, Tim Pohlemann, Antonius Pizanis, Swen Norrick, Emmanouil Lioadiks, David Engel

Saarland University, Faculty of Medicine, Deparment for Trauma, Hand and Reconstructive Surgery, Kirrbergerstr. 100, 66421, Homburg/Saar, Germany; Saarland University Medical Center, Department for Trauma, Hand and Reconstructive Surgery, Kirrbergerstr. 100, 66421, Homburg/Saar, Germany. Electronic address: tobias.fritz@uks.eu. · Saarland University, Faculty of Medicine, Deparment for Trauma, Hand and Reconstructive Surgery, Kirrbergerstr. 100, 66421, Homburg/Saar, Germany.

Key points

From the paper's conclusion
  • Fiber-Tape cerclage augmentation significantly enhances distraction resistance and preserves contact area compared to dual plating. Furthermore, this technique will potentially allow improved radiological visualization of the area of interest in the pubic symphysis after arthrodesis to analyze bone healing and osteointegration.

Abstract

BACKGROUND: Chronic anterior pelvic instability may require symphyseal arthrodesis when conservative treatment and isolated plating fail. Therefore, dual plating has been clinically used to improve stability. As a novel augmentation FiberTape® cerclage may provide and alternative stabilization method with comparable mechanical stability, providing radiolucency for improved analysis of bone healing. Therefore, we conducted a biomechanical study to compare the effects of dual plating to a plate/ FiberTape® cerclage construct. METHODS: Twenty-four synthetic pelvic models (n = 8 per group) were stabilized using symphyseal plating, dual plating, or plate-cerclage constructs. Cyclic axial loading (50-400 N) was applied in a unilateral stance and bilateral-leg stance biomechanical model. RESULTS: The plate-cerclage resulted in significantly higher compression force (131.5 ± 31.7 N) compared to plate-only fixation (88.2 ± 20.9 N; p = 0.019), corresponding to a 49% increase. Dual plating reduced contact area (p ≤ 0.049). No significant differences were observed in one-leg stance. CONCLUSION: Fiber-Tape cerclage augmentation significantly enhances distraction resistance and preserves contact area compared to dual plating. Furthermore, this technique will potentially allow improved radiological visualization of the area of interest in the pubic symphysis after arthrodesis to analyze bone healing and osteointegration.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers