Comparison of novel anterior fixation method, W-plate, versus superior parasymphyseal plating in anterior posterior compression type III pelvic ring injuries: a biomechanical study.
Brandi Ann Krieg, Todd Baldini, Christopher Orear, Keenan Masami Onodera, Nicholas Jerome Tucker, Adam Schlauch, Justin E Hellwinkel, Nathaniel Elliot Schaffer,
Department of Orthopedic Surgery, University of Colorado School of Medicine. · Department of Orthopedic Surgery, Denver Health Medical Center.
Key points
From the paper's conclusion- In a biomechanical analysis of induced APC-III pelvic ring injury, W-plating performed similarly to SPSP in stiffness, displacement, and LTF. W-plating for anterior pelvic ring fixation was found to be biomechanically similar to SPSP with potentially increased resistance to anterior ring displacement.
Abstract
OBJECTIVES: To compare stiffness, displacement, and load to failure (LTF) patterns of a novel anterior pelvic plating technique, W-plating, to superior parasymphyseal plating (SPSP) in anterior posterior compression type III (APC III) pelvic ring injuries in a biomechanical model. METHODS: Twelve fourth generation composite pelves with induced APC III injuries stabilized with posterior screw fixation and either W-plating or SPSP were stressed via the Instron machine under single leg stance (SLS) axial load and external rotation (ER) load. Motion was analyzed at the pubic symphysis (PS) and sacroiliac joints (SIJ). PS and SIJ displacement, rotation, and stiffness were compared between the W-plate and SPSP groups under simulated stressed conditions. T-test was used to compare means and significance was set at p<0.05. Unadjusted p-values were reported alongside effect sizes (Cohen's d) and 95% confidence intervals (CI) to provide additional context regarding the magnitude and potential biomechanical relevance of the observed differences. RESULTS: The W-plate demonstrated less inferior PS displacement than the SPSP under SLS with a large observed effect size (0.5 ± 0.4 mm vs 1.2 ± 0.5 mm, respectively, Cohen's d = 1.47). Average PS displacement also demonstrated a large effect size favoring the W-plate construct (Cohen's d = 1.07), though statistical significance was not reached (0.3 ± 0.3 mm vs 0.7 ± 0.3 mm respectively, p = 0.07). There were no significant differences in stiffness in the SLS for the W-plate versus SPSP (98.9 ± 11.2 vs 105 ± 6.6 N/mm, respectively, p = 0.6) or in ER (4.4 ± 2.2 v 3.1 ± 1.6, respectively, p = 0.3). When loaded to failure, the W-plate and the SPSP both failed at the posterior fixation with no significant difference between groups (1411± 221 v 1545 ± 225 N, respectively p = 0.4). CONCLUSIONS: In a biomechanical analysis of induced APC-III pelvic ring injury, W-plating performed similarly to SPSP in stiffness, displacement, and LTF. W-plating for anterior pelvic ring fixation was found to be biomechanically similar to SPSP with potentially increased resistance to anterior ring displacement. LEVEL OF EVIDENCE: V.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical Comparison of Three Fixation Constructs for Tile Type C1.2 Pelvic Ring Fractures: A Finite Element AnalysisLife · 2026
- A novel stiffness-distribution-controlled hybrid locking compression plate for osteoporotic midshaft clavicle fractures: a finite element study.BMC musculoskeletal disorders · 2026
- Axial Stiffness Advantage of Nail–Plate Fixation in 101 Patient-Specific Femora: A Finite Element StudyJournal of Clinical Medicine · 2026
- Salvage Management of Anterior-Posterior Compression (APC) Type II Pelvic Ring Injury Using Isolated Posterior Fixation After Anterior Implant Removal: A Case Report.Cureus · 2026
- Can posterolateral tibial plateau fractures be treated with an anterolateral approach? A biomechanical study on novel plating concepts in artificial tibia models.Injury · 2026
- Biomechanical and finite element analysis of a novel annular multi-axial locking plate for patellar fracture fixation.Journal of orthopaedic surgery and research · 2026