Mammography equipment characteristics and inter-site variability in quality control and radiation dose: an IAEA multi-country study.

Virginia Tsapaki, Elizabeth Keavey, Carolina Rabin, Maryam Al-Hashim, Amaal Al-Rasbi, Jorge Patricio Castillo Lopez, Paulo R Costa, Isabelle Fitton,

Dosimetry and Medical Radiation Physics Section, Human Health Division, Department of Nuclear Science and Applications, International Atomic Energy Agency, Vienna, Austria. Electronic address: v.tsapaki@iaea.org. · National Breast Screening Programme: Breastcheck, Dublin, Ireland. Electronic address: Elizabeth.Keavey@screeningservice.ie.

Key points

From the paper's conclusion
  • This multi-country IAEA study provides a descriptive characterization of mammography equipment and associated QC/dose measurements. The IAEA ATIA tool is suitable as is for constancy testing; provided protocols are followed-particularly regarding positioning to prevent the phantom's copper structure from interfering with the automatic exposure control (AEC). Cross-facility comparisons are limited by heterogeneous system designs, protocol choices, and phantom builds; thus, universal IQ reference levels are impractical. Manufacturer/model-specific ranges, clearer AEC guidance, raw image access, and strict phantom specifications are recommended.

Abstract

OBJECTIVES: This study is part of the International Atomic Energy Agency (IAEA) Coordinated Research Project (CRP) "Advanced Tools for Quality and Dosimetry of Digital Imaging in Radiology" which aimed to a) analyze trends in mammography equipment across a wide range of clinical settings and b) assess a recently developed IAEA automated quality control (QC) method. MATERIALS AND METHODS: IAEA CRP participating sites applied the IAEA QC tool (Automated Tool for Image Analysis (ATIA)) to capture image quality (IQ) and dose metrics, ensuring comparable assessments across diverse healthcare environments. A comprehensive dataset, from April 2022 to December 2024, was compiled and analyzed to achieve the aforementioned aims. RESULTS: Across 15 countries and 43 institutions, 67 mammography systems contributed 3384 data points for initial analysis. The sample included 9 vendors, 13 full field digital mammography models, plus 2 systems with computed radiography plates. Raw images (∼90%) supported consistent cross-site IQ metrics; for-presentation images showed wider variability due to vendor processing. CONCLUSION: This multi-country IAEA study provides a descriptive characterization of mammography equipment and associated QC/dose measurements. The IAEA ATIA tool is suitable as is for constancy testing; provided protocols are followed-particularly regarding positioning to prevent the phantom's copper structure from interfering with the automatic exposure control (AEC). Cross-facility comparisons are limited by heterogeneous system designs, protocol choices, and phantom builds; thus, universal IQ reference levels are impractical. Manufacturer/model-specific ranges, clearer AEC guidance, raw image access, and strict phantom specifications are recommended.