Technical performance evaluation of mobile x-ray systems for chest imaging.

N W Marshall, J Binst, H Verhoeven, T Stinglhamber, W Michiels, H Bosmans

Competentiecentrum Medische Stralingsfysica, Radiologie, UZ Leuven, Herestraat 49, 3000 Leuven, Belgium; Medical Physics and Quality Assessment, Katholieke Universiteit Leuven 3000 Leuven, Belgium. Electronic address: nicholas.marshall@uzleuven.be. · Competentiecentrum Medische Stralingsfysica, Radiologie, UZ Leuven, Herestraat 49, 3000 Leuven, Belgium.

Key points

From the paper's conclusion
  • Despite some variation between detectors, overall imaging performance for the systems was good. These data could act as reference values for medical physicists assessing x-ray devices for bedside imaging.

Abstract

PURPOSE: To present both quantitative and test object-based imaging performance measurements for the latest mobile digital radiography (DR) x-ray systems for bedside imaging. METHODS: Seven mobile DR systems with CsI-based detectors were assessed. The modulation transfer function (MTF) and detective quantum efficiency (DQE) were measured using the RQA8 beam quality at three detector incident air kerma levels (KB). A weighted polynomial model applied to the variance was used to estimate electronic, quantum and structure noise fractions versus KB. Technical image quality was assessed by imaging the Leeds TO12 contrast-detail phantom with 10 cm, 15 cm and 20 cm PMMA, using the x-ray technique factors specified by the applications specialists for the adult Chest AP examination. RESULTS: Electronic noise dominated below a mean value of 0.14 µGy, while quantum noise comprised 91.6% ,on average, of total image noise at 2.5 µGy. The spatial frequency for an MTF of 0.1 ranged from 3.2 mm-1 to 4.7 mm-1, with a mean of 3.8 mm-1. DQE at 0.5 mm-1, measured at the RQA8 quality and a target KB value of 2.5 µGy, ranged from 0.46 to 0.64 with a mean of 0.55. Mean nominal threshold contrast for the 2 mm diameter disc (±1 standard deviation) was 1.72% ± 0.23%, 3.66% ± 0.68% and 8.05% ± 2.78%, for 10 cm, 15 cm and 20 cm PMMA, respectively. CONCLUSION: Despite some variation between detectors, overall imaging performance for the systems was good. These data could act as reference values for medical physicists assessing x-ray devices for bedside imaging.

Read the paper