Technical performance evaluation of mobile x-ray systems for chest imaging.
N W Marshall, J Binst, H Verhoeven, T Stinglhamber, W Michiels, H Bosmans
Competentiecentrum Medische Stralingsfysica, Radiologie, UZ Leuven, Herestraat 49, 3000 Leuven, Belgium; Medical Physics and Quality Assessment, Katholieke Universiteit Leuven 3000 Leuven, Belgium. Electronic address: nicholas.marshall@uzleuven.be. · Competentiecentrum Medische Stralingsfysica, Radiologie, UZ Leuven, Herestraat 49, 3000 Leuven, Belgium.
Key points
From the paper's conclusion- Despite some variation between detectors, overall imaging performance for the systems was good. These data could act as reference values for medical physicists assessing x-ray devices for bedside imaging.
Abstract
PURPOSE: To present both quantitative and test object-based imaging performance measurements for the latest mobile digital radiography (DR) x-ray systems for bedside imaging. METHODS: Seven mobile DR systems with CsI-based detectors were assessed. The modulation transfer function (MTF) and detective quantum efficiency (DQE) were measured using the RQA8 beam quality at three detector incident air kerma levels (KB). A weighted polynomial model applied to the variance was used to estimate electronic, quantum and structure noise fractions versus KB. Technical image quality was assessed by imaging the Leeds TO12 contrast-detail phantom with 10 cm, 15 cm and 20 cm PMMA, using the x-ray technique factors specified by the applications specialists for the adult Chest AP examination. RESULTS: Electronic noise dominated below a mean value of 0.14 µGy, while quantum noise comprised 91.6% ,on average, of total image noise at 2.5 µGy. The spatial frequency for an MTF of 0.1 ranged from 3.2 mm-1 to 4.7 mm-1, with a mean of 3.8 mm-1. DQE at 0.5 mm-1, measured at the RQA8 quality and a target KB value of 2.5 µGy, ranged from 0.46 to 0.64 with a mean of 0.55. Mean nominal threshold contrast for the 2 mm diameter disc (±1 standard deviation) was 1.72% ± 0.23%, 3.66% ± 0.68% and 8.05% ± 2.78%, for 10 cm, 15 cm and 20 cm PMMA, respectively. CONCLUSION: Despite some variation between detectors, overall imaging performance for the systems was good. These data could act as reference values for medical physicists assessing x-ray devices for bedside imaging.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of X-ray image quality and scattered radiation among portable, mobile, and stationary X-ray machines.Radiation and environmental biophysics · 2026
- Mammography equipment characteristics and inter-site variability in quality control and radiation dose: an IAEA multi-country study.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Detective quantum efficiency (DQE)-oriented design of a transparent scintillator tandem x-ray/optical microscope for photon-starved 7.2-keV radiography at the National Ignition Facility (NIF).The Review of scientific instruments · 2026
- Large-area flat-panel detector versus conventional detector for full-spine digital radiography: a retrospective comparative study.Frontiers in medicine · 2026
- Model‐based image restoration targeting higher‐dose digital mammography: Impact on radiologists' detection of clustered microcalcificationsMedical Physics · 2026
- Benchmarking technical image quality in dual-energy contrast-enhanced mammography.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026