A National Audit of Mammography Systems Settings That May Affect the Output of Artificial Intelligence Software
Alistair Mackenzie, John Loveland, Leila Farshadi, Carlijn Roozemond, Ruben E. van Engen
Royal Surrey NHS Foundation Trust · Maidstone and Tunbridge Wells NHS Trust
Key points
From the paper's conclusion- In the longitudinal audit, there were between 2 and 6 software versions in concurrent use on the Hologic systems tested by MPD1.
Abstract
Background: Artificial Intelligence (AI) software in mammography is trained on a set of processed images and may be less effective when applied to images acquired on different systems or systems with different processing and/or acquisition settings. The aim of this work was to undertake a retrospective audit of a large number of mammography systems in the United Kingdom and identify the number of differences in image acquisition and processing factors. Methods: Images of the TORMAM phantom are acquired as part of the routine quality control programme. Data from the DICOM header of these images were extracted to provide a snapshot in time of the system configurations. A longitudinal audit of DICOM header data for all of the Hologic systems was tested by one medical physics department (MPD1) over 14 years. Results: We received results from 28 UK medical physics services for 498 systems. There were 7 different models of mammography systems, each with up to 7 different versions of acquisition workstation software. Each mammographic model had multiple image processing versions, including bespoke settings. The GE had two dose settings, while Siemens systems had a range of doses from 80% to 150% of the standard dose. In the longitudinal audit, there were between 2 and 6 software versions in concurrent use on the Hologic systems tested by MPD1. Conclusions: This study showed the heterogeneity of system setup across the UK in a single year, as well as changes to system setup over time. These differences may affect the outcomes of both AI and human readers. There are responsibilities on AI suppliers, mammography equipment manufacturers, breast-screening units, and medical physics services to ensure outcomes are not adversely affected by differences or changes in mammography equipment configurations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Subcategorisation of Data for AI Models in Healthcare: A Case Study in Mammography.Diagnostics (Basel, Switzerland) · 2026
- Mammography equipment characteristics and inter-site variability in quality control and radiation dose: an IAEA multi-country study.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Diagnostic Performance and Concordance of a BI-RADS-Output AI System in Mammography: A Retrospective Study of Radiologists-AI Collaboration.Journal of imaging informatics in medicine · 2026
- Investigating Artificial Intelligence Performance on Mammographic Cases in High- and Low-Resourced Countries.Asia-Pacific journal of clinical oncology · 2026
- Computational Harmonization of Digital Mammography: A Review of DICOM-Aware Preprocessing, Feature Reproducibility, and Cross-Vendor Domain ShiftAlgorithms · 2026
- Performance of a female-trained mammography AI system in male mammography across gynecomastia patternsFrontiers in Radiology · 2026