Benchmarking technical image quality in dual-energy contrast-enhanced mammography.
A Barcella, A Sarno, C Lenardi, L Brambilla, E Pignoli, T Giandini, N Paruccini, R Villa,
Department of Physics, University of Milan, Milano, Italy; INFN section of Milan, Milano, Italy. · Fondazione IRCCS Istituto Nazionale dei Tumori, Medical Physics Unit, Milano, Italy.
Key points
From the paper's conclusion- This study establishes typical values that could function as reference standards in QC programs. TCC0.5 enables quantitative comparison of residual glandular signal across CEM systems and may serve as a practical warning indicator in QC programs, identifying deviations from effective background suppression that warrant further investigation.
Abstract
PURPOSE: The aim of this study was to evaluate the technical image quality of Dual-energy Contrast-Enhanced Mammography (CEM) across different mammography systems and to establish typical values for quality control (QC) programs, which are currently not available. METHODS: 29 mammography systems (6 models, 5 vendors) in 23 hospitals were evaluated using the CIRS Model 022 phantom. The difference in mean pixel value (MPV) between iodine and background was measured at different iodine concentration and fitted with linear regression constrained to the origin. The TCC0.5 index, defined as the ratio between the signal difference of a 100% glandular plug and that of a 0.5 mg/cm2 iodine plug, was used toquantify the system's background subtraction performance. RESULTS: The analysis of system signal linearity as a function of iodine concentration provides typical values in terms of the slope (m) and the regression coefficient (R2) of the linear fits. TCC0.5 consistently reflected the visually perceived residual glandular signal across different systems. Three TCC0.5 ranges were identified to provide a practical classification of the observed residual glandular signal: negligible (|TCC0.5| < 0.15), perceptible (0.15 ≤ |TCC0.5| < 0.30), and marked residual signal (|TCC0.5| ≥ 0.30). Improved TCC0.5 values were observed for one system after the LE filter and software update. CONCLUSIONS: This study establishes typical values that could function as reference standards in QC programs. TCC0.5 enables quantitative comparison of residual glandular signal across CEM systems and may serve as a practical warning indicator in QC programs, identifying deviations from effective background suppression that warrant further investigation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Variability and bias in the calculation of iodine uptake with an adaptive background subtraction method in contrast-enhanced mammography.Physics in medicine and biology · 2026
- Diagnostic performance of the proposed DECT NI-RADS criteria for imaging surveillance of post-treatment head and neck squamous cell carcinoma: A preliminary study.The neuroradiology journal · 2026
- ROI-Based Quantitative Analysis in Contrast-Enhanced Mammography: A Feasibility Study for Reproducible Enhancement Assessment.Diagnostics (Basel, Switzerland) · 2026
- Mammography equipment characteristics and inter-site variability in quality control and radiation dose: an IAEA multi-country study.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Technical performance evaluation of mobile x-ray systems for chest imaging.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Feasibility of Contrast-Enhanced Spectral Mammography in Japan: A Pilot Study.JMA journal · 2026