Influence of anterior plate fixation on axial load-bearing capacity after single-level anterior cervical discectomy and fusion: A biomechanical cadaver study.

Maximilian Weber, Vincent Heck, Tamara Babasiz, Maximilian Lenz, Lars Peter Mueller, Peer Eysel, Sebastian Wegmann

University of Cologne, Faculty of Medicine and University Hospital, Center for Orthopedic and Trauma Surgery, Kerpener Str. 62, 50937 Cologne, Germany. Electronic address: maximilian.weber1@uk-koeln.de. · University of Cologne, Faculty of Medicine and University Hospital, Center for Orthopedic and Trauma Surgery, Kerpener Str. 62, 50937 Cologne, Germany.

Key points

From the paper's conclusion
  • Anterior plating was associated with higher maximum axial forces in this small cadaveric series. The findings suggest a possible contribution to axial load-bearing capacity but should be considered preliminary given the limited sample size and restricted loading protocol.

Abstract

BACKGROUND: Anterior cervical discectomy and fusion (ACDF) is widely used for degenerative and traumatic cervical spine pathologies. The contribution of anterior plate fixation to axial load-bearing capacity remains unclear. This cadaveric study compared stand-alone and plated single-level ACDF constructs under axial compression. METHODS: Five fresh-frozen human cervical spine specimens underwent ACDF at C3/C4 using either a stand-alone cage or additional anterior plate fixation. One specimen was excluded because of a technical measurement error, leaving two specimens per group. Constructs were subjected to monotonic axial compression programmed at 5 N/s. Maximum axial force recorded during testing (Fmax) was the primary outcome. FINDINGS: Both plated specimens reached higher Fmax values than the stand-alone constructs. Mean Fmax was 1746.1 ± 45.7 N with plating and 1563.3 ± 13.4 N without plating, corresponding to an 11.7% higher mean value. INTERPRETATION: Anterior plating was associated with higher maximum axial forces in this small cadaveric series. The findings suggest a possible contribution to axial load-bearing capacity but should be considered preliminary given the limited sample size and restricted loading protocol.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers