Analytical performance of CSF, pleural, and peritoneal fluids on the Siemens Atellica CH930.
Kwaku Twum, Steven W Cotten
Department of Pathology and Laboratory Medicine, Boston University Chobanian & Avedisian School of Medicine, Boston, MA, United States. Electronic address: Kwaku.twum@bmc.org. · Department of Pathology and Laboratory Medicine, University of North Carolina at Chapel Hill, Chapel Hill, NC, United States. Electronic address: steven.cotten@unchealth.unc.edu.
Key points
From the paper's conclusion- In conclusion, the Siemens Atellica CH930 platform can be successfully validated for body fluid testing; however, clinical laboratories may look to apply matrix-specific HIL thresholds rather than relying on conventional serum indices.
Abstract
OBJECTIVES: Biochemical analysis of body fluids is clinically essential and a requirement for compliance with Clinical Laboratory Improvement Amendments laws. This study evaluates the analytical performance specifications including hemolysis, icterus and lipemia (HIL) interference thresholds for 13 common chemistry analytes in cerebrospinal, pleural and peritoneal fluids on the Siemens Atellica CH930. DESIGN & METHODS: Residual fluid samples were used to evaluate precision, accuracy, and linearity across the analytical measurement range. Body fluid specific HIL interference studies were done using spiked hemolysate, commercial bilirubin ditaurate, and residual lipemic fluid. RESULTS: Inter-assay precision coefficient of variation (CV) was predominantly below 10% and aligned with the manufacturer's claims for standard fluids. Mixing studies also yielded acceptable mean percent recoveries ranging from 94% to 106% and linearity (R2 > 0.99) for most analytes, except lactate dehydrogenase with a lower recovery of 78%. Lastly, several analytes exceeded allowable error limits at lower HIL concentrations compared to standard serum claims. CONCLUSIONS: In conclusion, the Siemens Atellica CH930 platform can be successfully validated for body fluid testing; however, clinical laboratories may look to apply matrix-specific HIL thresholds rather than relying on conventional serum indices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-302 Body Fluids as Complex Analytical Metrices: Challenges in Validation and Analytical Performance of Automated Biochemistry AnalyzersClinical Chemistry · 2026
- Assessment of Hemolysis, Icterus, and Lipemia Interference in 14 Clinical Chemistry and 5 Immunoassay Parameters.Clinical laboratory · 2026
- Optimization of hemolysis index thresholds for biochemical and immunoassays.Bioanalysis · 2026
- Simulation Study on Preanalytical Variables in Coagulation Testing: Interference Effects and Threshold Determination of Hemolysis, Lipemia, Icterus and Insufficient Sample Volume.Clinical and applied thrombosis/hemostasis : official journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis · 2026
- B-025 Evaluating Preanalytical Strategies for body Fluids: Effects of Hyaluronidase Treatment and Filtration on Biochemical TestingClinical Chemistry · 2026
- Re-evaluating haemolysis rejection criteria for PT, aPTT, and D-dimer testing: evidence from paired clinical samples on the CN-3000 analyser.Advances in laboratory medicine · 2026