Re-evaluating haemolysis rejection criteria for PT, aPTT, and D-dimer testing: evidence from paired clinical samples on the CN-3000 analyser.
Jirapa Kaewkhruawan, Wararat Masalae, Surapat Tanticharoenkarn, Pitchayaporn Riyagoon, Waroonkarn Laiklang, Kwanlada Chaiwong, Suppakorn Wongkamchai, Phandee Watanaboonyongcharoen,
Division of Laboratory Medicine, King Chulalongkorn Memorial Hospital, Bangkok, Thailand. · Department of Laboratory Medicine, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
Key points
From the paper's conclusion- Mild to moderate spontaneous haemolysis does not produce clinically significant interference in PT, aPTT, or D-dimer. CN-3000 flagging behaviour is predictable and grade-dependent. Re-evaluation of automatic haemolysis rejection thresholds is warranted.
Abstract
OBJECTIVES: Haemolysis is the leading preanalytical cause of sample rejection in coagulation testing. This study prospectively evaluated the impact of spontaneous haemolysis on routine haemostasis assays by comparing paired haemolysed (H) and non-haemolysed (NH) clinical samples. Additionally, analyser flagging behaviour was characterised using an artificially haemolysed sample series. METHODS: Thirty nine paired of H and NH samples were analysed for prothrombin time (PT), activated partial thromboplastin time (aPTT), and D-dimer (DD) on the Sysmex CN-3000. Mean biases were compared against predefined critical differences (CD) derived from biological variation data. Analyser flagging was additionally assessed using five sample pools across five haemolysis concentration levels (H1-H5). RESULTS: When comparing H samples with NH samples, the mean bias was +0.07 s (+0.14 %) for PT (p=0.5979), +0.44 s (+0.66 %) for aPTT (p=0.8704), and +0.72 mg/L FEU (+10.63 %) for D-dimer (p=0.0199). D-dimer showed the highest relative bias but remained below the CD (64.7 %). Degree of haemolysis did not predict the magnitude of interference across the range of haemoglobin (Hb) concentrations studied. CN-3000 flagging grades were constant and correlated free Hb and irrespective of baseline sample PT/aPTT/D-dimer level. CONCLUSIONS: Mild to moderate spontaneous haemolysis does not produce clinically significant interference in PT, aPTT, or D-dimer. CN-3000 flagging behaviour is predictable and grade-dependent. Re-evaluation of automatic haemolysis rejection thresholds is warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Simulation Study on Preanalytical Variables in Coagulation Testing: Interference Effects and Threshold Determination of Hemolysis, Lipemia, Icterus and Insufficient Sample Volume.Clinical and applied thrombosis/hemostasis : official journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis · 2026
- Reduction in Sample Recollection Number Following the Implementation of Automated Pre-Analytical Interference Detection in Coagulation Testing.Annals of clinical biochemistry · 2026
- Assessment of Hemolysis, Icterus, and Lipemia Interference in 14 Clinical Chemistry and 5 Immunoassay Parameters.Clinical laboratory · 2026
- A-325 Optimizing Management of Hemolyzed Specimens: Enhancing Clinical Value and Reducing Troponin and Potassium RecollectionsClinical Chemistry · 2026
- Why do hemolysis interference studies produce different results? A proposal for standardized experimental design and reportingClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- Evaluation of Pre-Analytical Performance Using IFCC Quality Indicators, Six Sigma Metrics and Root Cause Analysis in a Biochemistry Laboratory-A Retrospective Study.Diagnostics (Basel, Switzerland) · 2026