Performance comparison of UniSprayTM and electrospray ionisation coupled to ion-mobility high-resolution mass spectrometer for screening of organic pollutants in water matrices.

Victoria Eriksson, Kalpana Munnuru Singamshetty, Thomas Malte Molnár Karlsson, Selina Tisler, Jan H Christensen, Johan Dahlén, Foon Yin Lai, Thanh Wang

Department of Physics, Chemistry, and Biology (IFM), Linköping University, SE-581 83, Linköping, Sweden. Electronic address: victoria.eriksson@liu.se. · Department of Thematic Studies (TEMA), Linköping University, SE-581 83, Linköping, Sweden. Electronic address: kalpana.munnuru.singamshetty@liu.se.

Key points

From the paper's conclusion
  • Overall, USI and ESI could provide complementary analytical response, as certain compounds are preferentially detected by one ion source over the other.

Abstract

Suspect and non-target screening (SNTS) by high-resolution mass spectrometry is widely used to characterise complex environmental mixtures, but detection depends strongly on ionisation efficiency and matrix effects. UniSprayTM ionisation (USI) is a relatively new ion source that applies a voltage to a metal rod positioned after a grounded spray capillary in contrast to electrospray ionisation (ESI), where the voltage is applied to the spray capillary. USI can increase signal intensity in targeted LC-MS, though its performance for SNTS of complex water matrices remains unclear. USI and ESI were compared using standard mixtures, evaluating signal intensity, reproducibility, linearity, limits of detection, adduct formation, and in-source fragmentation, and further assessed in wastewater and stormwater extracts via feature detection, suspect hits, matrix effects, and correlations with physicochemical parameters such as water-solubility (LogS), retention time, and drift time. USI produced higher signal intensities than ESI for most compounds. In standards, 48/58 (positive) and 47/60 (negative) showed higher peak areas with USI (2.6-3.0-fold). In environmental samples, 80-85% of features had higher peak areas (2.3-2.4-fold). USI detected more features in five of six extracts, especially low-intensity signals, without proportional increases in suspect annotations. In total, 94 (positive) and 54 (negative) features were annotated (levels 1-3), mainly shared between sources. The clearest advantage of USI was observed for low-abundance suspects in diluted stormwater, where several compounds were uniquely detected at lower enrichment factors and matrix effects were shown to be compound-specific. Correlations indicated that USI preferentially enhances compounds that are more water-soluble, earlier-eluting and have shorter drift times. Overall, USI and ESI could provide complementary analytical response, as certain compounds are preferentially detected by one ion source over the other. Combined use could improve chemical-space coverage and enhance the detection of environmentally relevant compounds in wastewater and stormwater screening.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Performance comparison of UniSprayTM and electrospray ionisation coupled to ion-mobility high-resolution mass spectrometer for screening of organic pollutants in water matrices — Science Skim