An Isotope-Induced Polarity Bias in Electrospray Ionization Mass Spectrometry.
Qi Ying Neo, Henry T P Annecke, Irena Papageorgiou, Axel Leppert, Katja Petzold, Kevin Pagel, Ali Miserez, Erik G Marklund,
Department of Microbiology, Tumor and Cell Biology, Karolinska Institutet, Solna 171 65, Sweden. · Department of Medical Biochemistry and Microbiology & Science for Life Laboratory & Center of Excellence for the Chemical Mechanisms of Life, Uppsala University, Uppsala Biomedical Centre, Uppsala 752 37, Sweden.
Key points
From the paper's conclusion- These findings reveal an unexpected solvent-dependent bias in negative-mode native MS measurements which can be exploited for experimental design.
Abstract
Heavy water (D2O) is commonly used as a solvent additive in electrospray ionization-mass spectrometry (ESI-MS) to probe biomolecular interactions under the assumption that isotopic substitution does not significantly affect the ionization process. However, D2O differs measurably from H2O in several physicochemical properties, which can influence ionization across biomolecular classes. Here, we systematically investigate how D2O influences native ESI across biomolecular classes and show that D2O introduces a pronounced, polarity-dependent detection bias which can suppress the signal of proteins and peptides in negative ion mode relative to other classes of biomolecules. Protein signal suppression is mitigated by reducing primary ESI droplet size or adding volatile ammonium salts, implicating altered droplet evolution and salt retention in D2O. Its magnitude is strongly dependent on analyte acidity: basic and moderately acidic proteins are most affected, whereas extremely negatively charged proteins, RNA, and polysaccharides remain largely unaffected. Signal loss also correlates with extensive sodium adduction and peak splitting rather than impaired charging. Exploiting this selectivity, we demonstrate controlled switching between peptide and RNA detection in a coacervate-based cellular delivery system. These findings reveal an unexpected solvent-dependent bias in negative-mode native MS measurements which can be exploited for experimental design.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Solvent-Isotope Effects in Biomolecular Phase Separation and Fibrillation of Disordered Proteins and Peptides.bioRxiv : the preprint server for biology · 2026
- Systematic assessment of different wash solvents for the analysis of small molecule metabolites using mass spectrometry imagingAnalytical and Bioanalytical Chemistry · 2026
- Revealing Hidden Variables in DESI-Based Spatial Metabolomics: Solvent Composition and Tissue Type as Critical Drivers.Journal of proteome research · 2026
- Influence of D/H Ratio in Water on Fluorescence and Biological Activity of Drug SolutionsHydrogen · 2026
- Performance comparison of UniSprayTM and electrospray ionisation coupled to ion-mobility high-resolution mass spectrometer for screening of organic pollutants in water matrices.Journal of chromatography. A · 2026
- Solvent Leveling Explains Supercharging in Electrospray Ionization Mass Spectrometry.Journal of the American Chemical Society · 2026