An Isotope-Induced Polarity Bias in Electrospray Ionization Mass Spectrometry.

Qi Ying Neo, Henry T P Annecke, Irena Papageorgiou, Axel Leppert, Katja Petzold, Kevin Pagel, Ali Miserez, Erik G Marklund,

Department of Microbiology, Tumor and Cell Biology, Karolinska Institutet, Solna 171 65, Sweden. · Department of Medical Biochemistry and Microbiology & Science for Life Laboratory & Center of Excellence for the Chemical Mechanisms of Life, Uppsala University, Uppsala Biomedical Centre, Uppsala 752 37, Sweden.

Key points

From the paper's conclusion
  • These findings reveal an unexpected solvent-dependent bias in negative-mode native MS measurements which can be exploited for experimental design.

Abstract

Heavy water (D2O) is commonly used as a solvent additive in electrospray ionization-mass spectrometry (ESI-MS) to probe biomolecular interactions under the assumption that isotopic substitution does not significantly affect the ionization process. However, D2O differs measurably from H2O in several physicochemical properties, which can influence ionization across biomolecular classes. Here, we systematically investigate how D2O influences native ESI across biomolecular classes and show that D2O introduces a pronounced, polarity-dependent detection bias which can suppress the signal of proteins and peptides in negative ion mode relative to other classes of biomolecules. Protein signal suppression is mitigated by reducing primary ESI droplet size or adding volatile ammonium salts, implicating altered droplet evolution and salt retention in D2O. Its magnitude is strongly dependent on analyte acidity: basic and moderately acidic proteins are most affected, whereas extremely negatively charged proteins, RNA, and polysaccharides remain largely unaffected. Signal loss also correlates with extensive sodium adduction and peak splitting rather than impaired charging. Exploiting this selectivity, we demonstrate controlled switching between peptide and RNA detection in a coacervate-based cellular delivery system. These findings reveal an unexpected solvent-dependent bias in negative-mode native MS measurements which can be exploited for experimental design.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

An Isotope-Induced Polarity Bias in Electrospray Ionization Mass Spectrometry — Science Skim